FDA Adverse Event Malfunction Summary report: N

ALLEN S.A.M. STIRRUP/TABLE CLAMP

MDR report key: 264508 · Received February 18, 2000

Report

Report Number
MW1018218
Event Type
Malfunction
Date Received
February 18, 2000
Date of Event
February 14, 2000
Report Date
February 15, 2000
Manufacturer
ALLEN MEDICAL SYSTEMS
Product Code
EYD
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE ALLEN STIRRUPS FELL DURING A VAGINAL PROCEDURE CAUSING THE PTS LEG TO FALL. "ANKLE WAS TWISTED" UPON INSPECTING ALLEN STIRRUPS NOTED THAT THE PORTION THAT GOES INTO THE HOLDER WAS WORN SMOOTH CAUSING THE STIRRUPS TO BE UNSTABLE. HOLDERS WERE SUPPLIED BY MFR OF THE ALLEN STIRRUPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEN S.A.M. STIRRUP/TABLE CLAMP * EYD ALLEN MEDICAL SYSTEMS ALLEN S.A.M. STIRRUP *
2 TABLE CLAMP TABLE CLAMP BWN ALLEN MEDICAL SYSTEMS 40017 *

Patients

Seq Age Sex Outcome Treatment
1 38 YR