FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2644211 · Received July 7, 2012

Report

Report Number
2050012-2012-01364
Event Type
Malfunction
Date Received
July 7, 2012
Date of Event
June 12, 2012
Report Date
June 12, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

EVALUATION RESULTS: WASH STATION. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT LEVEL 2 MULTIQUAL CONTROL HAS BOTTOMED OUT ON MULTIPLE CHEMISTRIES ON THE CARTRIDGE CHEMISTRY (CC) SIDE OF THE UNICEL DXC 600 SYNCHRON SYSTEM THE LAST TWO NIGHTS. CUSTOMER INDICATED THAT LEVEL 2 MULTIQUAL CONTROL WAS ACCEPTABLE WHEN REPEATED. CUSTOMER REPORTED THAT LEVEL 3 QUALITY CONTROL RAN DIRECTLY AFTER LEVEL 2 WAS ACCEPTABLE. CUSTOMER REPORTED THAT THE PROBLEM SEEMED TO HAVE STARTED AFTER THE NIGHT TECH REPLACED THE CAP PIERCER BLADE. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) CONFIRMED THE CAP PIERCER WAS LEAKING. THE FSE FOUND DEBRIS WAS PRESENT IN THE WASH STATION. THE FSE CLEARED THE DEBRIS FROM THE WASH STATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1