FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 2643592 · Received July 6, 2012

Report

Report Number
2124215-2012-07608
Event Type
Injury
Date Received
July 6, 2012
Date of Event
May 22, 2012
Report Date
August 16, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE TESTING WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE. MEASUREMENTS THROUGHOUT THIS TEST WAS WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD HAD DISLODGED INSIDE THE POCKET. THE LEAD WAS EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUITY IMPLANTABLE HF LEAD OJX CPI - DEL CARIBE 4591

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization| L| R 4591| 4479| E110| 0184| 4542| N119