FDA Adverse Event
Injury
Summary report: N
ETAC WALKER ONO
MDR report key: 2641811
·
Received June 28, 2012
Report
- Report Number
- 3008986539-2012-00023
- Event Type
- Injury
- Date Received
- June 28, 2012
- Date of Event
- April 22, 2009
- Report Date
- September 18, 2009
- Manufacturer
- ETAC SUPPLY CENTER AB
- Product Code
- ITJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 3008986539-1/27/12-001-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE WALKER'S LEFT REAR WHEEL CAME OFF. THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMING OFF HAD BEEN OVER-EXTENDED. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. THE WALKER WAS 9 YEARS OLD. (B)(4).
Description of Event or Problem · 1
THE USER WAS ALONE IN HIS ROOM WHEN THE RIGHT REAR WHEEL OF THE WALKER CAME OFF AND HE FELL. HE HAD A RIB FRACTURE AND BRUISES ON HIS BUTTOCKS AND RIGHT SIDE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETAC WALKER ONO | WALKER MECHANICAL - ITJ | ITJ | ETAC SUPPLY CENTER AB | 15040101-33 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |