FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 2641549
·
Received June 29, 2012
Report
- Report Number
- 2518422-2012-01343
- Event Type
- Malfunction
- Date Received
- June 29, 2012
- Date of Event
- January 30, 2012
- Report Date
- January 30, 2012
- Manufacturer
- RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER A FAILURE RELATED TO THE POWER SUPPLY WAS OBSERVED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. THE MALFUNCTION OF THE POWER SUPPLY WOULD RESULT IN THE FAILURE OF THE DEVICE TO OPERATE ON AC POWER OR TO CHARGE ITS INTERNAL BATTERY. THIS PARTICULAR MALFUNCTION FOR THE POWER SUPPLY HAS ONLY BEEN REPORTED ON DEVICES THAT ARE BEING SOLD INTERNATIONALLY FOR USE WITH 220 VOLT POWER SOURCES.
Description of Event or Problem · 1
THE MFR RECEIVED INFO ALLEGING A VENTILATOR REQUIRED SERVICE. THE DEVICE WAS NOT IN PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP | 1054096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|