FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 2641549 · Received June 29, 2012

Report

Report Number
2518422-2012-01343
Event Type
Malfunction
Date Received
June 29, 2012
Date of Event
January 30, 2012
Report Date
January 30, 2012
Manufacturer
RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER A FAILURE RELATED TO THE POWER SUPPLY WAS OBSERVED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. THE MALFUNCTION OF THE POWER SUPPLY WOULD RESULT IN THE FAILURE OF THE DEVICE TO OPERATE ON AC POWER OR TO CHARGE ITS INTERNAL BATTERY. THIS PARTICULAR MALFUNCTION FOR THE POWER SUPPLY HAS ONLY BEEN REPORTED ON DEVICES THAT ARE BEING SOLD INTERNATIONALLY FOR USE WITH 220 VOLT POWER SOURCES.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A VENTILATOR REQUIRED SERVICE. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP 1054096