FDA Adverse Event Malfunction Summary report: N

ETAC WALKER ONO

MDR report key: 2639513 · Received June 28, 2012

Report

Report Number
3008986539-2012-00034
Event Type
Malfunction
Date Received
June 28, 2012
Date of Event
April 14, 2011
Report Date
May 13, 2011
Manufacturer
ETAC SUPPLY CENTER AB
Product Code
ITJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
3008986539-1/27/12-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. THE WHEEL HAS A BEARING PLACED IN A HOUSING ON EACH SIDE OF THE WHEEL. THE INNER SIDE OF THE TIRE AND INNER HOUSING OF THE LEFT REAR WHEEL WERE DAMAGED. THE OUTER BEARING HAD A MARK FROM THE EDGE OF THE CIRCLIP. (B)(4).

Description of Event or Problem · 1

ON THE MORNING OF (B)(6) 2011, HOME CARE STAFF REPORTED THAT THE LEFT REAR WHEEL OF THE WALKER HAD FALLEN OFF WHEN THE USER ROSE. THE USER HAD FALLEN THE NIGHT BEFORE, ON (B)(6) 2011. WHILE DRAWING THE CURTAIN, HE STUMBLED AND FELL, AND THE WALKER LANDED ON TOP OF HIM. NO INJURY TO THE USER WAS REPORTED. A PHYSIOTHERAPIST VISITED THE USER ON (B)(6) 2011 TO REPLACE THE WALKER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETAC WALKER ONO WALKER MECHANICAL - ITJ ITJ ETAC SUPPLY CENTER AB 15040405-14 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK