FDA Adverse Event
Malfunction
Summary report: N
ETAC WALKER ONO
MDR report key: 2639513
·
Received June 28, 2012
Report
- Report Number
- 3008986539-2012-00034
- Event Type
- Malfunction
- Date Received
- June 28, 2012
- Date of Event
- April 14, 2011
- Report Date
- May 13, 2011
- Manufacturer
- ETAC SUPPLY CENTER AB
- Product Code
- ITJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 3008986539-1/27/12-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. THE WHEEL HAS A BEARING PLACED IN A HOUSING ON EACH SIDE OF THE WHEEL. THE INNER SIDE OF THE TIRE AND INNER HOUSING OF THE LEFT REAR WHEEL WERE DAMAGED. THE OUTER BEARING HAD A MARK FROM THE EDGE OF THE CIRCLIP. (B)(4).
Description of Event or Problem · 1
ON THE MORNING OF (B)(6) 2011, HOME CARE STAFF REPORTED THAT THE LEFT REAR WHEEL OF THE WALKER HAD FALLEN OFF WHEN THE USER ROSE. THE USER HAD FALLEN THE NIGHT BEFORE, ON (B)(6) 2011. WHILE DRAWING THE CURTAIN, HE STUMBLED AND FELL, AND THE WALKER LANDED ON TOP OF HIM. NO INJURY TO THE USER WAS REPORTED. A PHYSIOTHERAPIST VISITED THE USER ON (B)(6) 2011 TO REPLACE THE WALKER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETAC WALKER ONO | WALKER MECHANICAL - ITJ | ITJ | ETAC SUPPLY CENTER AB | 15040405-14 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |