FDA Adverse Event
Malfunction
Summary report: N
ETAC WALKER ONO
MDR report key: 2639486
·
Received June 28, 2012
Report
- Report Number
- 3008986539-2012-00022
- Event Type
- Malfunction
- Date Received
- June 28, 2012
- Date of Event
- June 30, 2009
- Report Date
- August 6, 2009
- Manufacturer
- ETAC SUPPLY CENTER AB
- Product Code
- ITJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 3008986539-1/27/12-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE WALKER'S LEFT REAR WHEEL HAD COME OFF. THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMING OFF HAD BEEN OVER-EXTENDED. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. ACCORDING TO THE CUSTOMER, THE WHEELS HAD NEVER BEEN REPLACED. HOWEVER, THE WHEEL THAT CAME OFF WAS LESS DUSTY AND LESS WORN THAN THE OTHER THREE WHEELS. IN ADDITION, THE CIRCLIPS OF THE THREE WHEELS THAT HAD NOT COME OFF WERE SLIGHTLY OVER-EXTENDED, AND THE WHEEL AXLES WERE LUBRICATED WITH A GREASE THAT ETAC DOES NOT USE. (B)(4).
Description of Event or Problem · 1
THE REPORTER RECEIVED AN ONO WALKER WITH A WHEEL THAT HAD FALLEN OFF. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETAC WALKER ONO | WALKER MECHANICAL - ITJ | ITJ | ETAC SUPPLY CENTER AB | 15040105-14 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |