FDA Adverse Event
Malfunction
Summary report: N
ETAC WALKER AVANT
MDR report key: 2639478
·
Received June 28, 2012
Report
- Report Number
- 3008986539-2012-00047
- Event Type
- Malfunction
- Date Received
- June 28, 2012
- Date of Event
- July 20, 2011
- Report Date
- October 4, 2011
- Manufacturer
- ETAC SUPPLY CENTER AB
- Product Code
- ITJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 3008986539-1/27/12-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMING OFF THE WALKER HAD BEEN OVER-EXTENDED. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPEC. ACCORDING TO THE CUSTOMER, THE WALKER HAD BEEN SERVICED SEVERAL TIMES. (B)(4).
Description of Event or Problem · 1
THE LEFT FRONT WHEEL CAME OFF THE WALKER AND THE USER FELL BUT WAS NOT INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETAC WALKER AVANT | WALKER MECHANICAL - ITJ | ITJ | ETAC SUPPLY CENTER AB | 15030103-15 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |