FDA Adverse Event Malfunction Summary report: N

ETAC WALKER AVANT

MDR report key: 2639478 · Received June 28, 2012

Report

Report Number
3008986539-2012-00047
Event Type
Malfunction
Date Received
June 28, 2012
Date of Event
July 20, 2011
Report Date
October 4, 2011
Manufacturer
ETAC SUPPLY CENTER AB
Product Code
ITJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
3008986539-1/27/12-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMING OFF THE WALKER HAD BEEN OVER-EXTENDED. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPEC. ACCORDING TO THE CUSTOMER, THE WALKER HAD BEEN SERVICED SEVERAL TIMES. (B)(4).

Description of Event or Problem · 1

THE LEFT FRONT WHEEL CAME OFF THE WALKER AND THE USER FELL BUT WAS NOT INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETAC WALKER AVANT WALKER MECHANICAL - ITJ ITJ ETAC SUPPLY CENTER AB 15030103-15 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK