FDA Adverse Event Malfunction Summary report: N

ETAC WALKER ONO

MDR report key: 2639468 · Received June 28, 2012

Report

Report Number
3008986539-2012-00036
Event Type
Malfunction
Date Received
June 28, 2012
Date of Event
June 23, 2011
Report Date
July 5, 2011
Manufacturer
ETAC SUPPLY CENTER AB
Product Code
ITJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
3008986539-1/27/12-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMINGOFF WAS LOOSE, BUT UNDAMAGED BEHIND THE WHEEL CAP. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. ACCORDING TO ETAC'S CUSTOMER, THE WALKER HAD BEEN RECONDITIONED WITHOUT WHEEL CHANGE. ETAC REF. NO. (B)(4).

Description of Event or Problem · 1

THE USER RECEIVED THE WALKER ON (B)(6) 2011 AND HAD HAD ANOTHER WALKER OF THE SAME MODEL SINCE (B)(6) 2010. THE USER WAS WALKING ON A WOODEN FLOOR WHEN THE RIGHT REAR WHEEL FELL OFF. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETAC WALKER ONO WALKER MECHANICAL - ITJ ITJ ETAC SUPPLY CENTER AB 15041201-14 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK