FDA Adverse Event
Malfunction
Summary report: N
ETAC WALKER ONO
MDR report key: 2639468
·
Received June 28, 2012
Report
- Report Number
- 3008986539-2012-00036
- Event Type
- Malfunction
- Date Received
- June 28, 2012
- Date of Event
- June 23, 2011
- Report Date
- July 5, 2011
- Manufacturer
- ETAC SUPPLY CENTER AB
- Product Code
- ITJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 3008986539-1/27/12-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE CIRCLIP THAT PREVENTS THE WHEEL FROM COMINGOFF WAS LOOSE, BUT UNDAMAGED BEHIND THE WHEEL CAP. THE WHEEL AXLES OF THE WALKER WERE ACCORDING TO SPECIFICATION. ACCORDING TO ETAC'S CUSTOMER, THE WALKER HAD BEEN RECONDITIONED WITHOUT WHEEL CHANGE. ETAC REF. NO. (B)(4).
Description of Event or Problem · 1
THE USER RECEIVED THE WALKER ON (B)(6) 2011 AND HAD HAD ANOTHER WALKER OF THE SAME MODEL SINCE (B)(6) 2010. THE USER WAS WALKING ON A WOODEN FLOOR WHEN THE RIGHT REAR WHEEL FELL OFF. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETAC WALKER ONO | WALKER MECHANICAL - ITJ | ITJ | ETAC SUPPLY CENTER AB | 15041201-14 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |