FDA Adverse Event Malfunction Summary report: N

ETAC WALKER ONO

MDR report key: 2639449 · Received June 28, 2012

Report

Report Number
3008986539-2012-00018
Event Type
Malfunction
Date Received
June 28, 2012
Date of Event
February 1, 2009
Report Date
February 16, 2009
Manufacturer
ETAC SUPPLY CENTER AB
Product Code
ITJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
3008986539-1/27/12-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE WHEEL CAME OFF DURING HANDLING OF THE WALKER. NO USER WAS INVOLVED. TO PREVENT THE WHEEL FROM COMING OFF, A CIRCLIP IS PLACED IN THE GROOVE OF THE WHEEL AXLE. THE WHEEL AXLES WERE ACCORDING TO SPECIFICATION. THE GROOVE OF THE RIGHT FRONT FORK AXLE WAS NOT ACCORDING TO SPECIFICATION. (B)(4).

Description of Event or Problem · 1

CUSTOMER COMPLAINT. THE WALKER HAD LOST THE RIGHT FRONT WHEEL. A SALES REPRESENTATIVE FORM ETAC COLLECTED THE WALKER AND TOOK IT TO THE MANUFACTURER'S SITE. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETAC WALKER ONO WALKER MECHANICAL - ITJ ITJ ETAC SUPPLY CENTER AB 15040105-14 NA

Patients

Seq Age Sex Outcome Treatment
1