FDA Adverse Event Malfunction Summary report: N

ARCHITECT CA 19-9XR

MDR report key: 2635002 · Received June 29, 2012

Report

Report Number
1415939-2012-01244
Event Type
Malfunction
Date Received
June 29, 2012
Report Date
May 31, 2012
Manufacturer
ABBOTT LABORATORIES
Product Code
NIG
PMA / PMN Number
K052000
Removal / Correction Number
1415939-05/30/12-002-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE ISSUE REVEALED THAT THE CONJUGATE COMPONENT OF THE ARCHITECT CA19-9 AFFECTED REAGENT LOTS CONTRIBUTED TO ELEVATED CA19-9 PATIENT SAMPLE CONCENTRATIONS. ALL AFFECTED LOTS SHARE THE SAME CONJUGATE COMPONENT. ABBOTT ISSUED A PRODUCT RECALL FOR THE SIX AFFECTED LOTS (08851M500, 08849M500, 10122M500, 08852M500, 08853M500, 10040M500) OF THE ARCHITECT CA19-9 REAGENT, INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DESTROY ANY REMAINING INVENTORY OF THE AFFECTED LOTS.

Description of Event or Problem · 1

THE CUSTOMER IS OBSERVING HIGHER THAN EXPECTED ARCHITECT CA 19-9XR ASSAY RESULTS WHEN USING REAGENT LOT 08851M500. THE CUSTOMER STATED ONE PATIENT SAMPLE GENERATED A FALSELY ELEVATED RESULT OF 18.1 U/ML USING ARCHITECT CA19-9XR LOT 08851M500. THE PATIENT GENERATED A CA 19-9 RESULT OF 4.44 U/ML USING LOT 12092M500. THE SAMPLE WAS THEN RUN ON A COBAS E601 ANALYZER AND A CA 19-9 RESULT OF 10.62 WAS GENERATED. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT CA 19-9XR NIG ABBOTT LABORATORIES 08851M500

Patients

Seq Age Sex Outcome Treatment
1 ARCH I2000SR LN 03M74-01 SN (B)(4)