FDA Adverse Event Malfunction Summary report: N

ARCHITECT CA 19-9XR

MDR report key: 2634994 · Received June 29, 2012

Report

Report Number
1415939-2012-01241
Event Type
Malfunction
Date Received
June 29, 2012
Report Date
March 21, 2012
Manufacturer
ABBOTT LABORATORIES
Product Code
NIG
PMA / PMN Number
K052000
Removal / Correction Number
1415939-05/30/12-002-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE ISSUE REVEALED THAT THE CONJUGATE COMPONENT OF THE ARCHITECT CA19-9 AFFECTED REAGENT LOTS CONTRIBUTED TO ELEVATED CA19-9 PATIENT SAMPLE CONCENTRATIONS. ALL AFFECTED LOTS SHARE THE SAME CONJUGATE COMPONENT. ABBOTT ISSUED A PRODUCT RECALL FOR THE SIX AFFECTED LOTS (08851M500, 08849M500, 10122M500, 08852M500, 08853M500, 10040M500) OF THE ARCHITECT CA19-9 REAGENT, INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DESTROY ANY REMAINING INVENTORY OF THE AFFECTED LOTS.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE CURRENT ARCHITECT CA19-9XR REAGENT LOT (08851M500) IN USE DEMONSTRATED A SLOPE OF 2.28 ON PATIENT RESULTS, WHILE A NEW LOT HAD A SLOPE OF 1.26. NO SPECIFIC PATIENT DATA WAS PROVIDED. THE CUSTOMER IS REQUESTING REPLACEMENT OF ALL BOXES OF CURRENT LOT. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT CA 19-9XR NIG ABBOTT LABORATORIES 08851M500

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT I2000SR 03M74-01 SN (B)(4)