ARCHITECT CA 19-9XR
Report
- Report Number
- 1415939-2012-01241
- Event Type
- Malfunction
- Date Received
- June 29, 2012
- Report Date
- March 21, 2012
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- NIG
- PMA / PMN Number
- K052000
- Removal / Correction Number
- 1415939-05/30/12-002-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
(B)(4). EVALUATION OF THE ISSUE REVEALED THAT THE CONJUGATE COMPONENT OF THE ARCHITECT CA19-9 AFFECTED REAGENT LOTS CONTRIBUTED TO ELEVATED CA19-9 PATIENT SAMPLE CONCENTRATIONS. ALL AFFECTED LOTS SHARE THE SAME CONJUGATE COMPONENT. ABBOTT ISSUED A PRODUCT RECALL FOR THE SIX AFFECTED LOTS (08851M500, 08849M500, 10122M500, 08852M500, 08853M500, 10040M500) OF THE ARCHITECT CA19-9 REAGENT, INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DESTROY ANY REMAINING INVENTORY OF THE AFFECTED LOTS.
THE CUSTOMER STATED THAT THE CURRENT ARCHITECT CA19-9XR REAGENT LOT (08851M500) IN USE DEMONSTRATED A SLOPE OF 2.28 ON PATIENT RESULTS, WHILE A NEW LOT HAD A SLOPE OF 1.26. NO SPECIFIC PATIENT DATA WAS PROVIDED. THE CUSTOMER IS REQUESTING REPLACEMENT OF ALL BOXES OF CURRENT LOT. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT CA 19-9XR | NIG | ABBOTT LABORATORIES | 08851M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT I2000SR 03M74-01 SN (B)(4) |