ACCESS
Report
- Report Number
- 6000001-2012-11981
- Event Type
- Malfunction
- Date Received
- June 29, 2012
- Date of Event
- May 29, 2012
- Report Date
- June 7, 2012
- Manufacturer
- BAXTER HEALTHCARE - AIBONITO
- Product Code
- FPA
- PMA / PMN Number
- K811078
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. A BATCH REVIEW CANNOT BE PERFORMED SINCE THERE WAS NO LOT NUMBER PROVIDED. IF THE SAMPLE IS RECEIVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED CONDITION OF Y-TYPE CATH EXT SET/MALE LUER ADAPTER IN WHICH CUSTOMER OBSERVED LEAK AT JOINT OF Y CONNECTION WAS CONFIRMED DURING SAMPLE EVALUATION. ACTUAL DEFECTIVE SAMPLE WAS AVAILABLE FOR EVALUATION. DURING VISUAL INSPECTION NO DEFECTS WERE OBSERVED. SLIDE CLAMP WAS OPENED AND CLOSED; FEMALE LUER ADAPTER WAS ATTACHED TO FLUID SOURCE, ELIMINATING THE AIR. THE MALE LUER ADAPTER WAS INSERTED TO THE VASCULAR ACCESS DEVICE USING FIRM PUSH AND TWIST MOVEMENT TO ENSURE THE CONNECTION. THE SAMPLE FAILED DURING PRIMING DURING PRESSURE TEST AT 8PSI. THE ASSIGNABLE ROOT CAUSE OF THE REPORTED CONDITION IS UNKNOWN. ADDITIONAL INFORMATION: THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
A CUSTOMER REPORTED TO BAXTER (B)(4) OF A Y-TYPE CATH EXT SET/MALE LUER ADAPTER IN WHICH CUSTOMER OBSERVED LEAK AT JOINT OF Y CONNECTION. THE REPORTED CONDITION OCCURRED DURING PATIENT USE. A PATIENT WAS INVOLVED, BUT THERE IS NO REPORT OF PATIENT/USER INJURY OR MEDICAL INTERVENTION WAS NEEDED IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - AIBONITO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |