FDA Adverse Event
Injury
Summary report: N
BD
MDR report key: 2634226
·
Received June 25, 2012
Report
- Report Number
- MW5025967
- Event Type
- Injury
- Date Received
- June 25, 2012
- Date of Event
- June 25, 2012
- Report Date
- June 25, 2012
- Manufacturer
- BECTON DISCKINSON AND CO
- Product Code
- KYX
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- WI, US
- Reporter Occupation
- PHARMACIST
Narratives
Description of Event or Problem · 1
BECTON DICKINSON "BD ORAL DISPENSING SYRINGE 1 ML CLEAR WITH TIP CAP" REF (B)(4) LOT# 2020057. BAGS OF 100. ROUGHLY HALF OF SEVERAL BAGS WE HAVE IN STOCK CONTAIN SYRINGES THAT WILL ACCEPT A NEEDLE. THEY ALL HAVE "FOR ORAL USE ONLY" STAMPED ON THEM. THE REST ARE THE CORRECT PRODUCT THAT HAS A LARGER TIP THAT WILL NOT ACCEPT A NEEDLE. WHEN I COMPARE THEM TO "BD" 1ML SYRINGE TUBERCULIN SLIP TIP" REF (B)(4) THEY ARE IDENTICAL EXCEPT FOR THE ORAL USE ONLY WARNING. THIS HAS A RISK OF AN ORAL PRODUCT BEING ADMINISTERED INTRAVENOUSLY THAT COULD CAUSE SEVERE HARM OR DEATH. I HAVE CONTACTED THE MFR, BUT HAVE NOT HEARD BACK FROM THEM YET. THERE HAS BEEN NO HARM TO A PT AND WE ARE SORTING OUT THE INCORRECT PRODUCT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD | BD ORAL DISPENSING SYRINGE | KYX | BECTON DISCKINSON AND CO | REF 305217 | 2020057 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |