FDA Adverse Event Injury Summary report: N

T-FLEX ASPHERIC

MDR report key: 2634028 · Received June 26, 2012

Report

Report Number
9611165-2012-00030
Event Type
Injury
Date Received
June 26, 2012
Date of Event
June 15, 2012
Report Date
June 15, 2012
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
HQL
PMA / PMN Number
P060011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE REFERENCE (B)(4) HAS BEEN ALLOCATED TO THIS COMPLAINT BY RAYNER INTRAOCULAR LENSES LTD. THE PHYSICIAN EXPLANTED THE T-FLEX ASPHERIC 623T INTRAOCULAR LENS. THIS INCISION HAD TO BE ENLARGED TO AID EXPLANTATION AND CONSEQUENTLY, SUTURING WAS NECESSARY. WHEN THE ORDER FOR THE T-FLEX ASPHERIC 623T SUBJECT TO THIS REPORT WAS PLACED THE HOSP INSTRUCTED RAYNER NOT TO PROVIDE A BACK-UP T-FLEX ASPHERIC 623T INTRAOCULAR LENS OF THE SAME POWER THEREFORE, A SECONDARY SURGICAL PROCEDURE HAS BEEN SCHEDULED. OUR REVIEW OF PRODUCTION RECORDS OF THE T-FLEX ASPHERIC (B)(4) CONFIRMS THAT ALL MFG AND QUALITY CHECKS WERE CONDUCTED SUCCESSFULLY. ALL LENSES RELEASED FROM THIS BATCH WERE WITHIN TOLERANCE AND SPEC. COMPLAINTS SINCE APRIL 2012, THE MONTH OF MANUFACTURE, WERE REVIEWED AND WE CAN CONFIRM THAT NO COMPLAINTS, OF ANY TYPE, HAVE BEEN RECEIVED AGAINST THE T-FLEX ASPHERIC (B)(4). THE T-FLEX ASPHERIC 623T INTRAOCULAR LENS IS NOT AVAILABLE IN THE UNITED STATES OF AMERICA. HOWEVER, THE T-FLEX ASPHERIC 623T INTRAOCULAR LENS HAPTICS ARE THE SAME AS THOSE ON THE C-FLEX 570C AND C-FLEX ASPHERIC 970C INTRAOCULAR LENSES. THE C-FLEX 570C AND C-FLEX ASPHERIC 970C ARE AVAILABLE IN THE USA.

Description of Event or Problem · 1

RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM THE SPANISH DIST OF AN INCIDENT THAT OCCURRED DURING THE USE OF A T-FLEX ASPHERIC 623T INTRAOCULAR LENS. THE DESCRIPTION OF THE EVENT IS AS FOLLOWS "AFTER PREPARING THE IOL IN THE INJECTOR, AND INSERTING THE IOL INTO THE LENS BAG, I REALIZED THAT THE BACK HAPTIC WAS BROKEN IN THE POINT IT IS JOINED TO THE OPTIC."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 T-FLEX ASPHERIC HQL - INTRAOCULAR LENS HQL RAYNER INTRAOCULAR LENSES LTD. 623T 042E36601

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention