FDA Adverse Event Injury Summary report: N

PKS OMNI

MDR report key: 2631033 · Received June 21, 2012

Report

Report Number
2183680-2012-00039
Event Type
Injury
Date Received
June 21, 2012
Date of Event
March 26, 2012
Report Date
May 24, 2012
Manufacturer
GYRUS MEDICAL INC
Product Code
GEI
PMA / PMN Number
K081766
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS CONFIRMED. DURING OUR INVESTIGATION, WE FOUND THAT THE HUB WAS FRACTURED. THE DEVICE WAS RETURNED WITH A FRACTURED HUB, THE FRACTURE SITE WAS UNDER THE HEAT SHRINK. THE HEAT SHRINK COVERS THE SHAFT AND THE BACK OR PROXIMAL END OF THE JAWS. UPON REMOVAL OF THE HEAT SHRINK, THERE WERE NO PIECES OF THE CERAMIC HUB OBSERVED ON THE INSIDE OF THE HEAT SHRINK. THE HUB FRACTURED JUST BEHIND THE CTR PIVOT RIVET AT THE THIN TO THICK TRANSITION AREA OF THE PART. WHEN THE TWO PARTS OF THE HUB WERE PLACED BACK TOGETHER, ALL PARTS ARE ACCOUNTED FOR. FAILURES SUCH AS THESE CAN BE ASSOCIATED WITH EXCESSIVE FORCE OR ROTATIONAL FORCES BEING APPLIED TO THE DISTAL OR WORKING END OF THE DEVICE. WE WILL MONITOR THE COMPLAINT DATABASE FOR FURTHER OCCURRENCES.

Description of Event or Problem · 1

IT WAS REPORTED TO GYRUSACMI THAT DURING A SURGICAL PROCEDURE, THE TIP OF TWO OMNI BROKE PROXIMAL TO THE JAW DURING SURGERY. NO PT HARM WAS REPORTED. (SEE MEDWATCH #2183680-2012-00033 FOR THE FIRST DEVICE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PKS OMNI PKS OMNI GEI GYRUS MEDICAL INC 970010PC 384308KD

Patients

Seq Age Sex Outcome Treatment
1 UNK