FDA Adverse Event
Malfunction
Summary report: N
DARCO HEADED DRILL BIT 6.5MM
MDR report key: 2630928
·
Received June 25, 2012
Report
- Report Number
- 8043862-2012-00008
- Event Type
- Malfunction
- Date Received
- June 25, 2012
- Report Date
- June 21, 2012
- Manufacturer
- AAP IMPLANTATE AG
- Product Code
- HTW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE BROKEN DRILL HAS BEEN CHECKED BY THE SUPPLIER IN DETAIL. NO DEVIATION HAS BEEN IDENTIFIED FROM THE SPECIFICATION. THE SUPPLIER ASSUMES THAT THE DRILL COLLIDED WITH SOMETHING MADE OF METAL DURING SURGERY WHICH COULD BE THE DRILL GUIDE OR ANOTHER IMPLANT THAT HAS BEEN IMPLANTED BEFORE. UNFORTUNATELY THE DISTRIBUTOR (B)(4) DID NOT PROVIDE ANY INFO.
Description of Event or Problem · 1
ALLEGEDLY THE DRILL BITS FAILED/BROKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DARCO HEADED DRILL BIT 6.5MM | DRILL | HTW | AAP IMPLANTATE AG | IW706522 | W002 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |