FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2630332 · Received June 27, 2012

Report

Report Number
2050012-2012-01349
Event Type
Malfunction
Date Received
June 27, 2012
Date of Event
June 4, 2012
Report Date
June 4, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
MZV
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

ON (B)(6) 2012, CUSTOMER REPORTED THAT THE DXC 600 PRO GENERATED A FALSE HIGH POTASSIUM (K) RESULT OF 5.9 MMOL/L ON 1 PATIENT SAMPLE. THE RESULT WAS NOT REPORTED OUT OF THE LAB. CUSTOMER STATED THAT BECAUSE THE RESULT DID NOT MATCH PATIENT HISTORY, IT WAS REPEATED. CUSTOMER STATED THAT UPON REPEAT THE INSTRUMENT SUPPRESSED THE RESULT (DID NOT GENERATE A RESULT). AFTER TROUBLESHOOTING CORRECTED THE ISSUE, THE SAMPLE WAS RUN AGAIN WITH A RESULT OF 4.7 MMOL/L, AND THIS RESULT WAS REPORTED. CUSTOMER STATED THAT AT THE TIME OF THE EVENT, THE INSTRUMENT SUPPRESSED ALL THE OTHER ION SELECTIVE ELECTRODE (ISE) RESULTS FOR THIS PATIENT SAMPLE AND TWO OTHER PATIENT SAMPLES. CUSTOMER STATED THAT THE K RESULT WAS THE ONLY RESULT THAT PRINTED OUT A NUMERICAL VALUE. CUSTOMER STATED THAT QUALITY CONTROL (QC) PRIOR TO THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES, BUT THE CUSTOMER DID NOT PROVIDE THE ACTUAL DATA. CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING, VIA PHONE. CUSTOMER STATED THAT THE ISE REFERENCE REAGENT WAS LOW, AND CTS INSTRUCTED THE CUSTOMER TO LOAD NEW ISE REFERENCE AND ISE BUFFER REAGENTS AND RECALIBRATE. CUSTOMER REPORTED THAT THIS ACTION RESOLVED THE ISSUE. NO FURTHER PROBLEMS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE MZV BECKMAN COULTER, INC. N/A

Patients

Seq Age Sex Outcome Treatment
1