TIGER 2 SELF-ADVANCING NASO JEJUNAL FEEDING TUBE
Report
- Report Number
- 1820334-2012-00345
- Event Type
- Injury
- Date Received
- June 20, 2012
- Date of Event
- May 8, 2012
- Report Date
- June 8, 2012
- Manufacturer
- COOK, INC.
- Product Code
- KNT
- PMA / PMN Number
- K043203
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4): VESSEL PERFORATION IS NOT SPECIFICALLY LISTED IN THE INSTRUCTIONS FOR USE. BOWEL EROSION IS LISTED IN THE INSTRUCTIONS FOR USE. PER INFORMATION SUPPLIED BY THE CUSTOMER, NO DEVICE WILL BE RETURNED TO ASSIST IN THE INVESTIGATION. THE PRODUCT IS SHIPPED WITH INSTRUCTIONS FOR USE WHICH DETAILS THE APPROPRIATE CONTRAINDICATIONS, PRECAUTIONS, AND PLACEMENT TECHNIQUES. SPECIFICATION REQUIRES 100% VERIFICATION THAT THE CORRECT TUBING HAS BEEN SUPPLIED, THE DISTAL END IS CLOSED AND ROUNDED AND THE CATHETER SURFACE IS NOT DAMAGED. WE ARE UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE, HOWEVER, IT IS POSSIBLE THAT THE PATIENT CONDITION CONTRIBUTED TO THE EVENT. WE WILL CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS. THE APPROPRIATE INTERNAL PERSONNEL HAVE BEEN NOTIFIED. PER THE CONCLUSION OF QUALITY ENGINEERING RISK ANALYSIS, NO FURTHER MITIGATING ACTION IS NECESSARY AT THIS TIME.
A FEMALE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2012 WITH LIVER ABSCESS. HER COAGULATION WAS SLIGHTLY ELEVATED TO PT-16.1, INR-1.25. THE TIGER 2 SELF-ADVANCING NASO JEJUNAL FEEDING TUBE WAS INSERTED ON (B)(6), COAGULATION NORMAL. ON (B)(6), THE PATIENT'S PT WAS 15.9 AND INR 1.23. ON (B)(6), THE PATIENT DEVELOPED A GI BLEED (MELENA), PT-17.0, INR-1.33. A GI TRACK SCOPE WAS PERFORMED ON (B)(6) RESULTING IN TRACKING AND EROSION FOUND ALONG THE COURSE OF THE FEEDING TUBE. INFORMATION WAS PROVIDED THAT THESE RESULTS ARE BELIEVED TO BE CAUSED BY THE TIGER 2 SELF-ADVANCING NASO JEJUNAL FEEDING TUBE. THE PATIENT STARTED ON A PROTON-PUMP INHIBITOR AND THE FEEDING TUBE WAS DISCONTINUED. THE PATIENT SUFFERED FROM GI BLEED (MELENA). THE PATIENT'S CONDITION IS UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TIGER 2 SELF-ADVANCING NASO JEJUNAL FEEDING TUBE | KNT TUBES, GASTROINTESTINAL | KNT | COOK, INC. | NA | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |