CHEMETRON TIRO OXYGEN OUTLET STATION
Report
- Report Number
- 1924066-2012-00005
- Event Type
- Other
- Date Received
- June 11, 2012
- Date of Event
- May 16, 2012
- Report Date
- June 11, 2012
- Manufacturer
- ALLIED HEALTHCARE PRODUCTS, INC.
- Product Code
- CAN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
TRIO OUTLET STATIONS WERE LAST MADE IN 1996. THEREFORE THE TRIO OUTLET STATION IS AT LEAST 16 YEARS OLD. THE TRIO OUTLET STATION WAS PLACED IN THE OPENING UPSIDE DOWN. THIS IS INDICATED BY THE LABELING, IN THE PICTURE, BEING UPSIDE DOWN ON THE FACEPLATE. THE FACEPLATE IS STICKING OUT BECAUSE THE KEYING PIN WOULD NOT LET IT LATCH PROPERLY. HOSPITAL VENTILATORS GENERALLY HAVE OXYGEN METERS AND LOW OXYGEN ALARMS. APPARENTLY THESE WERE NOT CHECKED AFTER ATTACHING THE PT TO THE VENTILATOR. THE TRIO OUTLET STATIONS WERE A UNDERWRITERS LABORATORIES LISTED PRODUCT AT THE TIME OF MANUFACTURING. THE UNIT INVOLVED HAS NOT BEEN RETURNED FOR EVALUATION.
IT WAS REPORTED THAT A PT WAS INTUBATED AND PLACED ON A VENTILATOR. WITHIN MINUTES HE BECAME CYANOTIC AND BRADYCARDIC AND CODE 89. CALLED PHYSICIAN. PT WAS BAGGED AND HE RETURNED TO BASELINE. NEW VENTILATOR WAS PLACED ON PT AND RECEIVED LOW OXYGEN ALARM. IT WAS IMMEDIATELY NOTICED THAT THE OXYGEN CONNECTION FACEPLATE WAS NOT SECURE AND WAS UPSIDE DOWN IN THE WALL. THIRD VENTILATOR WAS HOOKED UP TO A DIFFERENT OXYGEN FACEPLATE AND WE HAD NO FURTHER ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHEMETRON TIRO OXYGEN OUTLET STATION | OXYGEN OUTLET STATION | CAN | ALLIED HEALTHCARE PRODUCTS, INC. | 88-47-3601 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other |