FDA Adverse Event Other Summary report: N

SEGMENTED CYLINDER APPLICATOR SET

MDR report key: 2625752 · Received June 15, 2012

Report

Report Number
9612638-2012-00009
Event Type
Other
Date Received
June 15, 2012
Date of Event
November 18, 2011
Report Date
May 21, 2012
Manufacturer
VARIAN MEDICAL SYSTEMS, INC.
Product Code
JAQ
PMA / PMN Number
K983436
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

USER DID NOT FOLLOW INSTRUCTIONS FOR USE AS THEY REPORTED THAT THE CLAMPING NUT FOR THE GUIDE TUBE WAS NOT TIGHTEN ENOUGH TO HOLD THE FLEXIBLE PROBE IN POSITION. THEREFORE THE FLEXIBLE PROBE SLIPPED OUT OF THE VAGINAL CYLINDER, WHICH RESULTED IN THE MISPOSITIONING OF THE RADIATION SOURCE. USER DID NOT FOLLOW INSTRUCTIONS FOR USE AS THEY USED PT THIGHS CLOSED AROUND THE CYLINDER AND GUIDING TUBE TO SECURE THE DEVICE DURING TREATMENT POSITION, RATHER THAN ADD'L ACCESSORIES AS DESCRIBED IN THE INSTRUCTION FOR USE. NO FURTHER F/U OF THIS MDR IS EXPECTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THEY USED A SEGMENTED CYLINDER APPLICATOR SET FOR THE LAST FRACTION OF A BRACHYTHERAPY TREATMENT ON (B)(6) 2011. AT THE TIME OF TREATMENT NO ABNORMAL CIRCUMSTANCE WAS NOTED FOR THAT TREATMENT FRACTION DURING OR AFTER TREATMENT. DURING THE PT'S F/U APPOINTMENT ON (B)(6) 2011, THE PHYSICIAN NOTED A REACTION ON THE LABIA BUT WASN'T SURE WHAT IT WAS AND APPARENTLY DIDN'T FEEL THAT IT WAS A RADIATION INDUCED REACTION. ON (B)(6) 2012, THEY REC'D A CALL FROM ANOTHER PHYSICIAN (DIFFERENT CLINIC) REPORTING THAT THE PT IS EXPERIENCING LATE STAGE TISSUE NECROSIS TO BOTH RIGHT AND LEFT INNER THIGHS. USING A PHOTOGRAPH PROVIDED BY ANOTHER PHYSICIAN, THE CLINICAL TEAM DETERMINED THAT IT IS LIKELY THE PROBE SLIPPED APPROX 15 CM PRIOR TO DELIVERY OF THE RADIATION FOR THE LAST FRACTION. THE LENGTH OF THE SKIN REACTION WAS ESTIMATED TO BE APPROX 7 CM CORRESPONDING TO THE ACTIVE LENGTH DELIVERED 15 CM PROXIMAL TO THE INTENDED LOCATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEGMENTED CYLINDER APPLICATOR SET BRACHYTHERAPY APPLICATOR JAQ VARIAN MEDICAL SYSTEMS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other