SEGMENTED CYLINDER APPLICATOR SET
Report
- Report Number
- 9612638-2012-00009
- Event Type
- Other
- Date Received
- June 15, 2012
- Date of Event
- November 18, 2011
- Report Date
- May 21, 2012
- Manufacturer
- VARIAN MEDICAL SYSTEMS, INC.
- Product Code
- JAQ
- PMA / PMN Number
- K983436
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
USER DID NOT FOLLOW INSTRUCTIONS FOR USE AS THEY REPORTED THAT THE CLAMPING NUT FOR THE GUIDE TUBE WAS NOT TIGHTEN ENOUGH TO HOLD THE FLEXIBLE PROBE IN POSITION. THEREFORE THE FLEXIBLE PROBE SLIPPED OUT OF THE VAGINAL CYLINDER, WHICH RESULTED IN THE MISPOSITIONING OF THE RADIATION SOURCE. USER DID NOT FOLLOW INSTRUCTIONS FOR USE AS THEY USED PT THIGHS CLOSED AROUND THE CYLINDER AND GUIDING TUBE TO SECURE THE DEVICE DURING TREATMENT POSITION, RATHER THAN ADD'L ACCESSORIES AS DESCRIBED IN THE INSTRUCTION FOR USE. NO FURTHER F/U OF THIS MDR IS EXPECTED.
CUSTOMER REPORTED THAT THEY USED A SEGMENTED CYLINDER APPLICATOR SET FOR THE LAST FRACTION OF A BRACHYTHERAPY TREATMENT ON (B)(6) 2011. AT THE TIME OF TREATMENT NO ABNORMAL CIRCUMSTANCE WAS NOTED FOR THAT TREATMENT FRACTION DURING OR AFTER TREATMENT. DURING THE PT'S F/U APPOINTMENT ON (B)(6) 2011, THE PHYSICIAN NOTED A REACTION ON THE LABIA BUT WASN'T SURE WHAT IT WAS AND APPARENTLY DIDN'T FEEL THAT IT WAS A RADIATION INDUCED REACTION. ON (B)(6) 2012, THEY REC'D A CALL FROM ANOTHER PHYSICIAN (DIFFERENT CLINIC) REPORTING THAT THE PT IS EXPERIENCING LATE STAGE TISSUE NECROSIS TO BOTH RIGHT AND LEFT INNER THIGHS. USING A PHOTOGRAPH PROVIDED BY ANOTHER PHYSICIAN, THE CLINICAL TEAM DETERMINED THAT IT IS LIKELY THE PROBE SLIPPED APPROX 15 CM PRIOR TO DELIVERY OF THE RADIATION FOR THE LAST FRACTION. THE LENGTH OF THE SKIN REACTION WAS ESTIMATED TO BE APPROX 7 CM CORRESPONDING TO THE ACTIVE LENGTH DELIVERED 15 CM PROXIMAL TO THE INTENDED LOCATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEGMENTED CYLINDER APPLICATOR SET | BRACHYTHERAPY APPLICATOR | JAQ | VARIAN MEDICAL SYSTEMS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |