UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM
Report
- Report Number
- 2122870-2012-01498
- Event Type
- Malfunction
- Date Received
- June 21, 2012
- Date of Event
- May 22, 2012
- Report Date
- May 23, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- NBC
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER STATED THAT THE PATIENT'S SAMPLE WAS A SHORT DRAW FROM AN EDTA TUBE AND SPUN DOWN FOR 10 MINUTES AT 3200 RPMS. THE CUSTOMER ALSO INDICATED THAT ALL SYSTEM CHECKS PRIOR TO THE EVENT WERE WITHIN THE SPECIFIED PARAMETERS. QC RESULTS WERE ALSO WITHIN THE LABORATORY'S EXPECTED RANGE. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND DISCOVERED A FLUID LEAK IN THE SUBSTRATE MANIFOLD. THE FSE REPLACED THE SUBSTRATE MANIFOLD AND CLEANED THE WASH WHEEL. THE FSE ALSO PERFORMED PIPETTOR ALIGNMENTS AND PRIMED THE SYSTEM. VALIDATION TESTING WAS PERFORMED ON BOTH THE INSTRUMENT AND THE CTA (CLOSED TUBE ALIQUOTER), WHICH NO FURTHER ISSUES WERE NOTED. ALTHOUGH HARDWARE MAY HAVE BEEN A CONTRIBUTING FACTOR, PRE-ANALYTICAL ISSUES MAY ALSO HAVE CONTRIBUTED TO THE LOW BNP2 RESULT. A DEFINITIVE CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO OBTAINING AN ERRONEOUSLY LOW B-TYPE NATRIURETIC PEPTIDE (BNP2) RESULT OF 1 PG/ML GENERATED ON THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. THE CUSTOMER DECIDED TO REANALYZE THE SAMPLE DUE TO SOME REOCCURRING ISSUES WITH OTHER ERRONEOUS RESULTS, WHICH IS DOCUMENTED IN MDR 2122870-2012-01482. SUBSEQUENT TESTING PRODUCED A BNP2 RESULT OF 127 PG/NL. THE ERRONEOUS BNP2 RESULT WAS NOT REPORTED OUT OF THE LABORATORY. PATIENT TREATMENT WAS NOT AFFECTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | NBC | BECKMAN COULTER INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |