SPECTRA OPTIA COLLECT ST
Report
- Report Number
- 1722028-2012-00357
- Event Type
- Other
- Date Received
- June 12, 2012
- Date of Event
- May 10, 2012
- Report Date
- May 16, 2012
- Manufacturer
- TERUMO BCT
- Product Code
- LKN
- PMA / PMN Number
- K103090
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). INVESTIGATION EVAL AND CORRECTIVE ACTIONS ARE IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.
(B)(4). THIS REPORT IS BEING FILED TO PROVIDE ADDITIONAL INFORMATION. TWO DISPOSABLE SETS WERE RECEIVED. SET 1: USED/WET SET. THE TUBING WAS LOOSE FROM THE BROAD (LOOP) SIDE OF WHITE CELL CHAMBER. INSPECTION OF TUBE INDICATES THAT IT WAS LIKELY INSERTED 2/3 OF BOND LENGTH (I.E. THERE WAS A 1/3 SOCKET BOND GAP). POTENTIAL CAUSES: INADEQUATE INSERTION OF THE TUBE INTO THE BOND SOCKET, LACK OF SOLVENT, DELAY BETWEEN THE SOLVENT APPLICATION AND BONDING. SET 2: CLEAN/DRY SET: THE TUBING WAS LOOSE FROM BOTH SIDES OF THE WHITE CELL CHAMBER. INSPECTION OF THE TUBES INDICATES THAT THEY WERE LIKELY INSERTED 2/3 OF BOND LENGTH (I.E. THERE WAS A 1/3 SOCKET BOND GAP). POTENTIAL CAUSES: INADEQUATE INSERTION OF THE TUBE INTO THE BOND SOCKET , LACK OF SOLVENT, DELAY BETWEEN THE SOLVENT APPLICATION AND BONDING, AND/OR CUSTOMER TESTING THE SET BY PULLING ON THE TUBING. INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.
TERUMO BCT INTERNAL REFERENCE: (B)(4). A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THIS UNIT SHOWED NO IRREGULARITIES DURING MANUFACTURING THAT WERE RELEVANT TO THIS ISSUE. ROOT CAUSE: THE DISPOSABLE SETS WERE RETURNED FOR INVESTIGATION AND NO GENERAL MIS-ASSEMBLIES WERE NOTED. THE ROOT CAUSE IS UNDETERMINED AT THIS TIME. POSSIBLE ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING: BOND FAILURE OF COLLECT TUBING TO THE CHAMBER. WEAKENING OF THE BOND BY PULL-TESTING PRIOR TO USE.
THE CUSTOMER REPORTED THAT DURING LEUKAPHERESIS (USING OPTIA SEPARATOR), THERE WAS DISJUNCTION OF TUBING AT THE JUNCTION BETWEEN THE OUTLET OF THE CELLS COLLECTION CHAMBER AND THE COLLECTING TUBING. THERE WAS A LEAK ALARM. THE PROCEDURE WAS INTERRUPTED WITHOUT HARM TO THE PT. CONSEQUENCES: PREMATURE TERMINATION OF COLLECTION OF STEM CELLS, A SECOND LEUKAPHERESIS WAS NEEDED TO COMPLETE THE COLLECTION. TERUMO BCT IS AWAITING THE RETURN OF THE DISPOSABLE SET. THIS REPORT IS BEING FILED DUE TO THE CUSTOMER FILLING A SAE REPORT WITH THEIR LOCAL COMPETENT AUTHORITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRA OPTIA COLLECT ST | SEPARATOR, AUTOMATED, BLOOD CELL AND P1 | LKN | TERUMO BCT | 12T3127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |