TERUFLEX DOUBLE CPD/OPTISOL 450ML
Report
- Report Number
- 1722028-2012-00312
- Event Type
- Other
- Date Received
- June 13, 2012
- Date of Event
- May 17, 2012
- Report Date
- May 18, 2012
- Manufacturer
- TERUMO BCT
- Product Code
- KSR
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS REPORT IS BEING FILED TO PROVIDE ADDITIONAL INFORMATION. INVESTIGATION: THE MANUFACTURING RECORDS FROM BOTH THE (B)(6) FACTORIES WERE CHECKED FOR PROBLEMS RELATING TO THIS EVENT. NO ABNORMALITIES WERE FOUND. THE NEEDLE GUARDS OF 4 RETENTION SAMPLES FROM THE REFERENCED LOT NUMBER WERE EXAMINED. NO ABNORMALITIES OR DEFORMATIONS WERE FOUND. ALSO, IT WAS CONFIRMED THAT THE NEEDLES RETRACTED FULLY INTO THE NEEDLE GUARDS. THE NEEDLES DID NOT COME OUT OF THE NEEDLE GUARD WHEN APPLYING FORCE TO THE TUBE AT THE BASE OF THE HUB IN THE DIRECTION OF THE EXPOSED NEEDLE TIP. ROOT CAUSE: THE DEVICE WAS NOT AVAILABLE FOR ROOT CAUSE ANALYSIS. SIMILAR INCIDENTS THAT HAVE BEEN REPORTED IN THE PAST, HAVE BEEN THE RESULT OF THE USER'S LACK OF EXPERIENCE WITH THE DEVICE. IT CANNOT BE RULED OUT THAT THE INCIDENT OCCURRED BY THE OPERATOR'S INCORRECT TECHNIQUE DURING USE.
(B)(4). INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.
THE CUSTOMER STATED THAT UPON OBSERVATION OF THE NEEDLE SET LODGED IN THE SAFETY DEVICE, IT DID APPEAR THAT THE NEEDLE WAS SOMEWHAT SLANTED TOWARD THE RIGHT SIDE. THERE ARE 2 RIDGES THAT RUN THROUGH THE CENTER/TOP OF THE INSIDE OF THE SAFETY DEVICE. THE NEEDLE TIP ITSELF ENDED UP ALIGNED WITH THE RIGHT HAND RIDGE AS YOU LOOK INTO THE SAFETY DEVICE. NORMALLY THE NEEDLE LINES UP IN THE CENTER OF THE SAFETY DEVICE.
THE CUSTOMER REPORTED THAT AN EMPLOYEE GOT A NEEDLESTICK. SHE HELD THE NEEDLEGUARD STATIONARY WITH HER LEFT FOREFINGER AND THUMB. SHE POSITIONED NEEDLE HUB INTO THE SAFETY DEVICE. AS SHE PULLED ON THE TUBING TO WITHDRAW THE NEEDLE AND PULL IT BACK INTO THE NEEDLE GUARD, THE NEEDLE/HUB BECAME PARTIALLY LODGED IN THE SAFETY DEVICE TO PULL OUT OF HER GRASP, RECOIL, AND SUBSEQUENTLY STUCK HER LEFT FINGER. EMPLOYEE WAS SEEN IN EMERGENCY ROOM AS PART OF OUR OCCUPATIONAL EXPOSURE PROTOCOL. THE AGE AND WEIGHT OF THE EMPLOYEE PLACED AT RISK ARE UNKNOWN. DONOR INFORMATION: (B)(6), FEMALE. THE DISPOSABLE SET IS UNAVAILABLE FOR RETURN BECAUSE THE CUSTOMER DISCARDED IT. THIS REPORT IS BEING FILED DUE TO A NEEDLESTICK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUFLEX DOUBLE CPD/OPTISOL 450ML | BLOOD BAG | KSR | TERUMO BCT | 111021GC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |