ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00407
- Event Type
- Malfunction
- Date Received
- June 19, 2012
- Date of Event
- May 24, 2012
- Report Date
- May 28, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: ONLY THE DRYLINE TUBE OF THE COMPLAINT RT340 ADULT BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A HOLE ABOUT 10CM AWAY FROM THE CONNECTOR OF THE RETURNED DRYLINE TUBE. A LOT CHECK REVEALED THREE OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120127. CONCLUSION: IT WAS IDENTIFIED THAT DAMAGE TO THE RT340 DRYLINE TUBE COULD HAVE BEEN CAUSED DURING THE MANUFACTURING PROCESS, ALTHOUGH IT IS NOT A BATCH RELATED ISSUE. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY 10 MINUTES AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. IT WAS ALSO REPORTED THAT A PINHOLE WAS FOUND IN THE DRYLINE TUBE. THIS WAS OBSERVED PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT340 | 120127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |