FDA Adverse Event
Injury
Summary report: N
BARRETT V2 I/A TIP, STRAIGHT 0.3MM PORT
MDR report key: 2621607
·
Received June 14, 2012
Report
- Report Number
- 3019924-2012-00024
- Event Type
- Injury
- Date Received
- June 14, 2012
- Date of Event
- May 17, 2012
- Report Date
- June 13, 2012
- Manufacturer
- MICROSURGICAL TECHNOLOGY, INC.
- Product Code
- HQE
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED UNIT WAS CLEAN AND HAD NO APPARENT DAMAGE WHEN INSPECTED WITH THE NAKED EYE. INSPECTION AT 40X DEMONSTRATED THAT THE PORT IS WELL SHAPED WITH NO DEFECTS. THE REMAINDER OF THE TIP IS IN VERY GOOD CONDITION AND MEETS FINAL ACCEPTANCE CRITERIA. THERE ARE SEVERAL SMALL SMOOTH SPOTS IN THE SAND BLASTED SURFACE BUT THERE ARE NO HOLES OR ROUGHENED DAMAGE.
Description of Event or Problem · 1
AN (B)(4) SALES REP REPORTED A SERIOUS PT INJURY WHEN A SURGEON WAS POLISHING A POSTERIOR CAPSULE WITH A BARRETT V2 IRRIGATION/ASPIRATION TIP, BRT-1033-B, STRAIGHT, CAPSULE POLISH, 0.3 MM PORT. THE SURGEON WAS REQUIRED TO PERFORM AN ANTERIOR VITRECTOMY, PLACE AN IMPLANTED IOL IN THE SULCUS, AND CLOSE THE PARACENTESIS WITH SUTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BARRETT V2 I/A TIP, STRAIGHT 0.3MM PORT | IRRIGATION/ASPIRATION TIP | HQE | MICROSURGICAL TECHNOLOGY, INC. | BRT-1033-B | 045993 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |