FDA Adverse Event Injury Summary report: N

STEALTHSTATION TREATMENT GUIDANCE PLATFORM

MDR report key: 262131 · Received February 14, 2000

Report

Report Number
1723170-2000-00001
Event Type
Injury
Date Received
February 14, 2000
Date of Event
December 20, 1999
Report Date
February 9, 2000
Manufacturer
MEDTRONIC SURGICAL NAVIGATION
Product Code
HAW
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

IN 1999, A DOCTOR ATTEMPTED TO USE THE STEALTHSTATION TREATMENT GUIDANCE PLATFORM AS AN AID IN REMOVING A CRANIAL LESION. THE DOCTOR PERFORMED A FIRST REGISTRATION, THEN REMOVED THE SKIN AND BONE FLAPS. AT THIS POINT, IT WAS NOTICED THAT THE CRANIAL ARC HAD MOVED. IT APPEARS THAT THE ARC WAS NOT TIGHTENED SUFFICIENTLY. SINCE THE STEALTHSTATION SYSTEM REQUIRES THAT THE ARC REMAIN RIGID, THE DOCTOR COULD NOT PROCEED WITH THE ORIGINAL REGISTRATION. THE DOCTOR THEN PROCEEDED TO CLOSE THE SKIN AND BONE FLAPS, AND PERFORMED A SECOND REGISTRATION. AFTER DR PERFORMED THE SECOND REGISTRATION AND A SECOND LOCALIZATION, THE DOCTOR PROCEEDED WITH THE CASE. THE MEAN FIDUCIAL ERROR (MFE) WAS 3.3 MM, WITH A PREDICTED ACCURACY OF 2.1 MM. THE LESION WAS THOUGHT TO BE APPROX 2 CM UNDERNEATH THE DURA MATER AND APPROX 1.5 CM IN DIAMETER, HOWEVER, THE DOCTOR HAD DIFFICULTY LOCATING THE LESION. WHEN THE CASE WAS COMPLETED, THE DOCTOR REQUESTED AN ADDITIONAL SCAN AND REALIZED THAT THE LESION WAS STILL PRESENT. THE DOCTOR ALSO REALIZED THAT DURING THE COURSE OF SURGERY, THE MOTOR STRIP WAS HIT. THE PT IS NOW UNI-LATERAL HEMIPLEGIC ON THE RIGHT SIDE OF BODY (THE ENTIRE RIGHT SIDE OF THE BODY IS PARALYZED).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEALTHSTATION TREATMENT GUIDANCE PLATFORM IMAGE GUIDED STEREOTACTIC DEVICE HAW MEDTRONIC SURGICAL NAVIGATION STEALTHSTATION SYSTEM NA

Patients

Seq Age Sex Outcome Treatment
1 40 YR Disability