OLYMPUS TRACHEAL INTRUBATION FIBERSCOPE
Report
- Report Number
- 8010047-2012-00184
- Event Type
- Other
- Date Received
- June 7, 2012
- Date of Event
- May 8, 2012
- Report Date
- May 9, 2012
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- EOQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN MORE INFO REGARDING THE REPORT AND WAS INFORMED THAT THE LOANER DEVICE DID NOT FIT THE 37 FR (12.3 MM) ETT, AND AS A RESULT THE PT PROCEDURE WAS CONVERTED TO AN OPEN SURGERY. HOWEVER, A REVIEW OF THE ENDOSCOPE SPECIFICATION SHOWED THAT THE DISTAL END OUTER DIAMETER OF THE SCOPE IS 5.1 MM WHICH IS MUCH SMALLER THAN THE ETT THAT THE USER CLAIMED TO USE WITH THE DEVICE; THUS THE SCOPE SHOULD HAVE FIT PERFECTLY INTO THE ETT. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE EVALUATION REVEALED A MINOR DENT ON THE INSERTION TUBE, AND THE IMAGE TILT WAS SLIGHTLY OFF. THE DEVICE PASSED THE LEAK TEST. THIS DEVICE WILL BE SERVICED AND RETURNED TO STOCK. THE EXACT CAUSE OF THE USER'S EXPERIENCE COULD NOT BE DETERMINED BUT USER MISHANDLING COULD NOT BE RULED OUT AS A CONTRIBUTORY FACTOR. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN A EXCESS OF CAUTION.
THE USER FACILITY REPORTED THAT A LOANER DEVICE WAS USED ON A PROCEDURE TO REPAIR ATRIAL SEPTAL DEFECT; HOWEVER, THE DEVICE DID NOT FIT INTO A 37 FRENCH ENDOTRACHEAL TUBE (EIT) MADE BY (B)(4). THE PT WAS SAID TO HAVE BEEN SEDATED, AS A RESULT AN OPEN SURGERY (STERNOTOMY) WAS PERFORMED ON THE PT TO PLACE THE BRONCHO CAP. THE PROCEDURE WAS COMPLETED AND THERE WAS NO PT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS TRACHEAL INTRUBATION FIBERSCOPE | INTUBATION FIBERSCOPE | EOQ | OLYMPUS MEDICAL SYSTEMS CORPORATION | LF-TP | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |