FDA Adverse Event Injury Summary report: N

ASSURE 4 BLOOD GLUCOSE SYSTEM

MDR report key: 2618374 · Received June 15, 2012

Report

Report Number
1832816-2012-00058
Event Type
Injury
Date Received
June 15, 2012
Date of Event
June 1, 2012
Report Date
June 1, 2012
Manufacturer
APEX BIOTECHNOLOGY CORP.
Product Code
NBW
PMA / PMN Number
K070088
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE METER INVOLVED IN INCIDENT WAS RETURNED, BUT THE TEST STRIPS WERE NOT RETURNED. THE RETURNED METER HAS COSMETIC DAMAGE, HOWEVER THIS DID NOT AFFECT PRODUCT PERFORMANCE. THE RETURNED METER WAS EVALUATED WITH RETENTION SAMPLES OF THE SAME LOT OF TEST STRIPS INVOLVED IN THE INCIDENT AND PERFORMED TO SPECIFICATION. NO PERFORMANCE FAILURE DETECTED.

Description of Event or Problem · 1

CALLER INDICATED THE ASSURE 4 METER WAS GIVING VARIABLE READINGS. AT 7:30 AM TOOK READING ON METER AND THE RESULTS WERE 67 MG/DL. AT 7:31 AM WITH SAME METER AND NEW TEST STRIP THE RESULTS WERE 289 MG/DL. NURSE LISA ADMINISTERED INSULIN TO THE RESIDENT. UNABLE TO DETERMINE ANY USER ERROR. CALLER STATED THAT THE RESIDENT IS FEELING FINE AT THIS TIME. FACILITY DOES NOT HAVE CONTROLS, I HAVE URGED THEM TO GET SOME PLUS EXTRA FOR EMERGENCY SITUATIONS, AND TO KEEP THIS ITEM STOCKED AT ALL TIMES. REPLACING PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASSURE 4 BLOOD GLUCOSE SYSTEM BLOOD GLUCOSE TEST SYSTEM NBW APEX BIOTECHNOLOGY CORP. 560001 SS054P

Patients

Seq Age Sex Outcome Treatment
1 70 YR Life Threatening