ALTAIRE
Report
- Report Number
- 8030405-2012-00004
- Event Type
- Malfunction
- Date Received
- June 11, 2012
- Date of Event
- May 10, 2012
- Report Date
- May 11, 2012
- Manufacturer
- HITACHI MEDICAL CORP.
- Product Code
- LNH
- PMA / PMN Number
- K050602
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE ALTAIRE IS A 0.7 TESLA OPEN MAGNET. HITACHI FIELD SERVICE INVESTIGATED THE COMPLAINT AND FOUND THAT A TRIMMER CAPACITOR HAD FAILED IN THE TRANSMITTER CIRCUIT. THE FAILED CAPACITOR WAS THE SOURCE OF THE HEAT FELT BY THE PT. THE CAPACITOR WAS REPLACED AND AFTER TESTING CONFIRMED PROPER OPERATION, THE UNIT WAS RETURNED TO SERVICE. THE COMPONENT IN QUESTION IS UNDER THE TRANSMITTER COVER, WHICH IS THE SURFACE WHERE THE PT'S ARM MADE CONTACT. THE EXACT TEMPERATURE OF THE HOT SPOT WAS NOT MEASURED, BUT HITACHI DETERMINED THAT THE MELTING POINT OF THE PAD USED IS ABOVE THE 41 DEGREE C LIMIT SET FOR PT CONTACT SURFACES PER IEC/UL 60601-1. THERE WAS NO VISIBLE DAMAGE ON THE TOP SURFACE OF THE COVER DUE TO HEAT. HITACHI HAS NOT COMPLETED THE ENGINEERING ANALYSIS OF THE CAPACITOR TO UNDERSTAND THE FAILURE MODE AND THE CHANCE OF REOCCURRENCE. NORMALLY THE EXTREMITIES REMAIN ON THE PT TABLE SURFACE, WHICH IS 80CM IN WIDTH. THE TABLETOP SLIDES OVER THE TRANSMITTER COVER AS IT ENTERS THE MAGNET. IF THE PT'S ARMS ARE UNRESTRAINED, THEY CAN MOVE OFF THE TABLE AND ON TO THE TRANSMITTER COVER. HITACHI USER TRAINING INSTRUCTS TECHNOLOGISTS TO PROVIDE PADS TO PREVENT DIRECT CONTACT WITH THE TRANSMITTER COVER DUE TO THE POTENTIAL FOR RF THERMAL INJURY FROM CLOSE PROXIMITY TO THE TRANSMIT COIL, A RISK COMMON TO ALL MRI SYSTEMS.
ON (B)(6) 2012, A PT WAS BEING SCANNED ON THE HITACHI ALTAIRE MRI SYSTEM. DURING THE EXAM, THE PT MOVED THEIR ARM OFF THE SIDE OF THE PT TABLE ONTO THE COVER OF THE MAGNET. THE PT REPORTED THE AREA WAS HOT TO THE TOUCH, SO THE TECHNOLOGIST PLACED AN ACCESSORY PAD UNDER THE PT'S ARM AND CONTINUED THE EXAM. AFTER THE EXAM WAS COMPLETED, WHEN THE TECHNOLOGIST REMOVED THE PAD FROM THE COVER, THERE WAS A SPOT WHERE THE PAD SURFACE WAS MELTED (APPROX 2-3CM IN DIAMETER). THIS PT WAS THE LAST OF THE DAY. THE SITE PLACED A SERVICE CALL THE NEXT MORNING AND WAS TOLD NOT TO USE THE SYSTEM UNTIL THE ISSUE WAS RESOLVED. THE PT DID NOT SUSTAIN ANY INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALTAIRE | MAGNETIC RESONANCE DIAGNOSTIC DEVICE | LNH | HITACHI MEDICAL CORP. | ALTAIRE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |