FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2616959 · Received June 14, 2012

Report

Report Number
2517506-2012-00109
Event Type
Malfunction
Date Received
June 14, 2012
Date of Event
May 14, 2012
Report Date
May 21, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
JFY
PMA / PMN Number
K090330
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE IMPRECISE CREATININE RESULT IS USER ERROR. THE ACCOUNT SET THE DIGITS OF PRECISION SETTING TO ONE WHICH PROVIDED RESULTS IN INTEGER VALUES ONLY RATHER THAN THE PROPER THREE DIGITS OF PRECISION. THE LACK OF SIGNIFICANT DIGITS CAUSE RESULTS TO BE ROUNDED WITH RESULTING IMPRECISION AT LOW ANALYTE VALUES. THE SIEMENS HEALTHCARE DIAGNOSTICS INC. TECHNICAL SERVICE CENTER REPRESENTATIVE INSTRUCTED THE ACCOUNT TO CORRECT THE DIGITS OF PRECISION SETTING. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A GROUP OF IMPRECISE CREATININE RESULTS WAS OBTAINED ON PATIENT AND QC SAMPLES. RESULTS WERE WAS REPORTED TO THE PHYSICIAN. RESULTS WERE REPORTED IN WHOLE NUMBER VALUES ONLY. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED AS A RESULT OF THE IMPRECISE CREATININE RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE IMPRECISE CREATININE RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM ENZYMATIC CREATININE FLEX® REAGENT CARTRIDGE JFY SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW 12033AE

Patients

Seq Age Sex Outcome Treatment
1