PJ COMFORT ELECTRIC BREAST PUMP
MDR report key: 2616535
·
Received June 8, 2012
Report
- Report Number
- 3002297831-2012-00001
- Event Type
- Malfunction
- Date Received
- June 8, 2012
- Date of Event
- May 9, 2012
- Report Date
- May 9, 2012
- Manufacturer
- LIMERICK INC
- Product Code
- HGX
- PMA / PMN Number
- K051926
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION: POWER CORD PROBLEM CAN BE RELATED TO THE OPERATOR'S USE AND TECHNIQUE OF BUNDLING THE POWER ADAPTER AND ELECTRICAL CORDS FOR CARRYING OR STORAGE. HOWEVER, CURRENT POWER CORD INSULATION IS ONLY UL/CSA SPT-1 RATED. DESIGN CONSIDERATIONS: ELECTRICAL CORD WITH SPT-2 INSULATION RATING WILL PROVIDE GREATER INSULATION THICKNESS TO MITIGATE THE RUPTURING OF ELECTRICAL CORDS. CHANGING TO POWER ADAPTERS THAT PLUG DIRECTLY INTO WALL OUTLETS (I.E. WITHOUT POWER CORD) SHOULD BE A CONSIDERATION.
Description of Event or Problem · 1
ELECTRICAL SHORT IN DEVICE POWER CORD, RESULTING IN ELECTRICAL SPARKING AND SINGEING CONSUMERS BLOUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PJ COMFORT ELECTRIC BREAST PUMP | BREAST PUMP | HGX | LIMERICK INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening |