PJ COMFORT ELECTRIC BREAST PUMP

MDR report key: 2616535 · Received June 8, 2012

Report

Report Number
3002297831-2012-00001
Event Type
Malfunction
Date Received
June 8, 2012
Date of Event
May 9, 2012
Report Date
May 9, 2012
Manufacturer
LIMERICK INC
Product Code
HGX
PMA / PMN Number
K051926
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: POWER CORD PROBLEM CAN BE RELATED TO THE OPERATOR'S USE AND TECHNIQUE OF BUNDLING THE POWER ADAPTER AND ELECTRICAL CORDS FOR CARRYING OR STORAGE. HOWEVER, CURRENT POWER CORD INSULATION IS ONLY UL/CSA SPT-1 RATED. DESIGN CONSIDERATIONS: ELECTRICAL CORD WITH SPT-2 INSULATION RATING WILL PROVIDE GREATER INSULATION THICKNESS TO MITIGATE THE RUPTURING OF ELECTRICAL CORDS. CHANGING TO POWER ADAPTERS THAT PLUG DIRECTLY INTO WALL OUTLETS (I.E. WITHOUT POWER CORD) SHOULD BE A CONSIDERATION.

Description of Event or Problem · 1

ELECTRICAL SHORT IN DEVICE POWER CORD, RESULTING IN ELECTRICAL SPARKING AND SINGEING CONSUMERS BLOUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PJ COMFORT ELECTRIC BREAST PUMP BREAST PUMP HGX LIMERICK INC

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening