SPRINT QUATTRO SECURE
MDR report key: 2610545
·
Received June 11, 2012
Report
- Report Number
- 2649622-2012-07313
- Event Type
- Injury
- Date Received
- June 11, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE LEAD WAS NOT RETURNED, HOWEVER WE DID RECEIVE PERFORMANCE EVALUATION AND HAVE ANALYZED THAT DATA. PATIENT ALERTS FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(4) 2012 03:00:13 AND (B)(4) 2012 03:00:10. DAILY PACE IMPEDANCE TREND DATA SHOWS AN ABRUPT INCREASE FOR MAX VENTRICULAR PACE BI IMPEDANCE = 475 TO 4047 OHMS PEAK BETWEEN (B)(4) 2012 AND (B)(4) 2012.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED NOISE IN THE SENSING CIRCUIT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |