FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 SLIDEMAKER

MDR report key: 2609627 · Received June 11, 2012

Report

Report Number
1061932-2012-01759
Event Type
Malfunction
Date Received
June 11, 2012
Date of Event
May 14, 2012
Report Date
May 14, 2012
Manufacturer
BECKMAN COULTER INC.
Product Code
GKJ
PMA / PMN Number
CLASS1EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE THE DAY OF THE EVENT AND FOUND A HOLE IN THE TUBING THROUGH PINCH VALVE VL1 ON THE SLIDEMAKER SAMPLE ACCESS MODULE (SAM), WHICH IS HOUSED UNDER THE COVER OF THE LH 750 ANALYTIC STATION. FSE REPLACED THE TUBING WHICH RESOLVED THE LEAK. DURING THIS SERVICE, FSE INADVERTENTLY TOUCHED THE TUBING THROUGH THE ADJACENT PINCH VALVE VL2 AND THE LINE POPPED OFF. JUST AFTER THE FSE LEFT, THE CUSTOMER REPORTED THE SUBSEQUENT LEAK; THE FSE WAS PAGED AND RETURNED THE SAME DAY TO REPAIR THE LEAK. THE CAUSE OF THE LEAK WAS A HOLE IN TUBING AND THE SUBSEQUENT LEAK WAS CAUSED BECAUSE TUBING POPPED OFF DUE TO FSE ERROR. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT CLEAR LIQUID WAS LEAKING FROM THE TUBING CONNECTING THE COULTER LH 750 HEMATOLOGY ANALYZER TO THEIR COULTER LH 750 SLIDEMAKER. A FEW DROPS (APPROXIMATELY 0.05 CC OF THE FLUID WAS ON THE COUNTER BETWEEN THE INSTRUMENTS. IN ADDITION, A FEW DROPS APPROXIMATELY 0.2 CC) WAS OBSERVED UNDER THE LASER. THE SLIDEMAKER STOPPED MAKING SLIDES DURING LEAK AND THE LEAK CEASED WHEN THE OPERATOR DISABLED THE SOFTWARE CONNECTION TO THE SLIDEMAKER. THERE IS AN EXPOSURE CONTROL PLAN IN PLACE AT THE FACILITY. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LAB COAT. NO INJURY OR EXPOSURE WAS REPORTED. THERE WAS NO IMPACT TO PATIENT SAMPLES OR RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 750 SLIDEMAKER AUTOMATED SLIDE SPINNER GKJ BECKMAN COULTER INC. LH 750 SLIDEMAKER NA