FDA Adverse Event Injury Summary report: N

ENTRUST AT

MDR report key: 2608975 · Received June 11, 2012

Report

Report Number
6000144-2012-03289
Event Type
Injury
Date Received
June 11, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
Z-1172-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED NORMAL DEPLETION THAT MEETS 80% OF EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) AND THERE MAY HAVE BEEN FASTER THAN NORMAL BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD