FDA Adverse Event Malfunction Summary report: N

GETINGE

MDR report key: 2601770 · Received October 2, 2009

Report

Report Number
9616031-2009-00001
Event Type
Malfunction
Date Received
October 2, 2009
Date of Event
August 5, 2009
Report Date
August 5, 2009
Manufacturer
GETINGE DISINFECTION AB
Product Code
MDZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

DURING THE INSTALLATION OF A WASTER DISINFECTOR AND AFTER APPLYING POWER TO THE WASHER THE SERVICE REP NOTICED SMOKE AT THE GROUND CONNECTION. THE SERVICE REP OBSERVED THAT THE GROUND WIRE INSULATION HAD BEGUN TO MELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GETINGE WASHER DISINFECTOR MDZ GETINGE DISINFECTION AB 9125 NA

Patients

Seq Age Sex Outcome Treatment
1 NA