FDA Adverse Event Malfunction Summary report: N

SYBGO PLAZA

MDR report key: 2598713 · Received May 18, 2012

Report

Report Number
2240869-2012-00039
Event Type
Malfunction
Date Received
May 18, 2012
Date of Event
April 19, 2012
Report Date
April 19, 2012
Manufacturer
SIEMENS , AG.
Product Code
IZH
PMA / PMN Number
K042095
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A NEW UI (UPDATE INSTRUCTION) WILL BE DISTRIBUTED TO ALL AFFECTED CUSTOMERS. WITH THIS UI, ALL AFFECTED DATA WILL BE FOUND AND DELETE PROTECTION WILL BE SET ON THE DATA TO PREVENT DATA LOSS. FOR ALL NEW INSTALLATIONS, THIS UI WILL BE INSTALLED AS WELL. WITH THE NEW SOFTWARE, VA20D, THE DATA LOSS IS PREVENTED IN THE SOFTWARE. ADD'L MEASURES WILL BE TAKEN AS WELL; A TOOL (SET DELETE PROTECTION TOOL) WILL BE RELEASED TO FIND AFFECTED DATA AND SET THE DELETE PROTECTION FOR SUCH DATA. THIS TOOL SHALL BE ADDED TO THE MAINTENANCE TASK AT THE SITES AND HAS TO RUN ONCE A DAY. THIS WILL BE RELEASED AS A SEPARATE UPDATE INSTRUCTION (UI SY129/12/S). AS AN IMMEDIATE MEASURE TO PREVENT DATA LOSS, UPDATE OF STORED PROCEDURES IN DATABASE WILL BE DONE TO "DELETE PROTECT" AFFECTED DATA AND DOUBLE-CHECK BEFORE CHANGE OF "ARCHIVE STATUS" (IN SOFTWARE VD20D); FOR ALL NEW INSTALLATIONS, THE UI SY129/12/S HAS TO BE INSTALLED IN ADDITION TO THE NEW VD20D SOFTWARE; ARCHIVE NOW TOOL WILL BE ENHANCED TO ARCHIVE THE AFFECTED SERIES. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT IN THE SYNGO PLAZA ARCHIVE ONLINE STS (SHORT-TERM STORAGE SYSTEM) THERE ARE STUDY ENTRIES WITH ARCHIVED SYMBOL, BUT NO ENTRY IN THE ARCHIVE DB (NO LINK TO THE FILE IN THE LTS - LONG TERM STORAGE). IF THE STUDY IS DELETED IN THE ONLINE STS, THE PICTURES ARE LOST. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT. THIS EVENT OCCURRED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYBGO PLAZA SYSTEM, IMAGE PROCESSING, RADIOLOGICAL IZH SIEMENS , AG. 10592457 NA

Patients

Seq Age Sex Outcome Treatment
1 Other