FDA Adverse Event
Malfunction
Summary report: N
PENGUIN NUTRITIONAL WARMER
MDR report key: 2598442
·
Received May 24, 2012
Report
- Report Number
- 2598442
- Event Type
- Malfunction
- Date Received
- May 24, 2012
- Date of Event
- May 8, 2012
- Report Date
- May 8, 2012
- Manufacturer
- NEONATAL PRODUCT GROUP, INC.DBA CRECHE INNOVATION
- Product Code
- FPF
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- NY, US
Narratives
Description of Event or Problem · 1
AC (ALTERNATING CURRENT) POWER CORD WAS WIRED INCORRECTLY ON THE NEW UNIT. IT WAS FOUND THAT THE GROUND AND NEUTRAL WIRES WERE REVERSED.======================MANUFACTURER RESPONSE FOR BREAST MILK WARMER, PENGUIN NUTRITIONAL WARMER (PER SITE REPORTER).======================THE MANUFACTURER WAS ADVISED OF THE FINDING. THEY OFFERED TO REPLACE ALL POWER CORDS.OUT OF 26 NEW UNITS ONLY 1 POWER CORD WAS FOUND DEFECTIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PENGUIN NUTRITIONAL WARMER | BREAST MILK WARMER | FPF | NEONATAL PRODUCT GROUP, INC.DBA CRECHE INNOVATION | PNW-001-1.0 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |