FDA Adverse Event Malfunction Summary report: N

PENGUIN NUTRITIONAL WARMER

MDR report key: 2598442 · Received May 24, 2012

Report

Report Number
2598442
Event Type
Malfunction
Date Received
May 24, 2012
Date of Event
May 8, 2012
Report Date
May 8, 2012
Manufacturer
NEONATAL PRODUCT GROUP, INC.DBA CRECHE INNOVATION
Product Code
FPF
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NY, US

Narratives

Description of Event or Problem · 1

AC (ALTERNATING CURRENT) POWER CORD WAS WIRED INCORRECTLY ON THE NEW UNIT. IT WAS FOUND THAT THE GROUND AND NEUTRAL WIRES WERE REVERSED.======================MANUFACTURER RESPONSE FOR BREAST MILK WARMER, PENGUIN NUTRITIONAL WARMER (PER SITE REPORTER).======================THE MANUFACTURER WAS ADVISED OF THE FINDING. THEY OFFERED TO REPLACE ALL POWER CORDS.OUT OF 26 NEW UNITS ONLY 1 POWER CORD WAS FOUND DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PENGUIN NUTRITIONAL WARMER BREAST MILK WARMER FPF NEONATAL PRODUCT GROUP, INC.DBA CRECHE INNOVATION PNW-001-1.0 *

Patients

Seq Age Sex Outcome Treatment
1 *