FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXC 800 SYNCHRON SYSTEM
MDR report key: 2596384
·
Received June 1, 2012
Report
- Report Number
- 2050012-2012-01191
- Event Type
- Malfunction
- Date Received
- June 1, 2012
- Date of Event
- May 7, 2012
- Report Date
- May 7, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
CUSTOMER REPORTED TO BECKMAN COULTER, INC (BEC) THAT THE UNICEL DXC 800 SYNCHRON SYSTEM GENERATED ERRONEOUS ELECTROLYTES RESULTS. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE FLOWCELL, CARBON BRIDGE, ALL ELECTRODES AND MODULAR CHEMISTRY SAMPLE PROBE. THE FSE ALSO REBUILT THE ELECTROLYTE INJECTION CUP AND CHECKED ALIGNMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |