PLASMABLADE TISSUE DISSECTION DEVICE
Report
- Report Number
- 3007069406-2012-00094
- Event Type
- Other
- Date Received
- May 22, 2012
- Date of Event
- April 23, 2012
- Report Date
- April 23, 2012
- Manufacturer
- MEDTRONIC ADVANCED ENERGY LLC
- Product Code
- GEI
- PMA / PMN Number
- K083415
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
RESULT - VISUAL INSPECTION OF THE RETURNED DEVICE NOTED A WHITE ASH AT THE DISTAL TIP OF THE HOUSING. THE HOUSING APPEARED TO BE BURNT AS PER THE REPORTED EVENT. A REVIEW OF THE LOT HISTORY RECORD (LHR) FOR THIS LOT NUMBER FOUND THAT THE DEVICE MET ASSEMBLY AND PRODUCT SPECS, NO ANOMALIES WERE FOUND DURING MFG. CONCLUSION - BENCH TOP INVESTIGATION FOUND THAT THE SILICONE MATERIAL IN THE TONSIL TIP CAN WITHSTAND THE HIGH TEMPERATURES SEEN DURING NORMAL USE. WHEN EXPOSED TO TEMPERATURES IN EXCESS OF 400 DEGREES CELSIUS, IT WILL BURN, LEAVING A WHITE ASH RESIDUE BUT WILL NOT GENERATE ANY FLAME. IT WAS DETERMINED THAT IN A HIGH OXYGEN ATMOSPHERE, AN ACTIVE TONSIL TIP CAN IGNITE THE OXYGEN AND BURN THE TONSIL TIP HOUSING. IT WAS NOT POSSIBLE TO REPRODUCE THE BURNT TONSIL TIP HOUSING UNDER NORMAL ATMOSPHERIC CONDITIONS. END OF REPORT.
IT WAS REPORTED THAT THE PHYSICIAN OBSERVED A FLAME AROUND THE PLASTIC TIP OF THE BLADE. NO PT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLASMABLADE TISSUE DISSECTION DEVICE | GEI | MEDTRONIC ADVANCED ENERGY LLC | PLASMABLADE TNA | 52132 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | NOT AVAILABLE |