FDA Adverse Event Other Summary report: N

PLASMABLADE TISSUE DISSECTION DEVICE

MDR report key: 2593903 · Received May 23, 2012

Report

Report Number
3007069406-2012-00093
Event Type
Other
Date Received
May 23, 2012
Date of Event
April 19, 2012
Report Date
April 24, 2012
Manufacturer
MEDTRONIC ADVANCED ENERGY LLC
Product Code
GEI
PMA / PMN Number
K083415
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESULTS: VISUAL INSPECTION OF THE RETURNED DEVICE NOTED A WHITE ASH AT THE DISTAL TIP OF THE HOUSING. THE HOUSING APPEARED TO BE BURNT AS PER THE REPORTED EVENT. A REVIEW OF THE LOT HISTORY RECORD (LHR) FOR THIS LOT NUMBER FOUND THAT THE DEVICE MET ASSEMBLY AND PRODUCT SPECIFICATIONS, NO ANOMALIES WERE FOUND DURING MANUFACTURING. CONCLUSIONS: BENCH TOP INVESTIGATION FOUND THAT THE SILICONE MATERIAL IN THE TONSIL TIP CAN WITHSTAND THE HIGH TEMPERATURES SEEN DURING NORMAL USE. WHEN EXPOSED TO TEMPERATURES IN EXCESS OF 400 DEGREES C, IT WILL BURN, LEAVING A WHITE ASH RESIDUE BUT WILL NOT GENERATE ANY FLAME. IT WAS DETERMINED THAT IN A HIGH OXYGEN ATMOSPHERE, AN ACTIVE TONSIL TIP CAN IGNITE THE OXYGEN AND BURN THE TONSIL TIP HOUSING. IT WAS NOT POSSIBLE TO REPRODUCE THE BURNT TONSIL TIP HOUSING UNDER NORMAL ATMOSPHERIC CONDITIONS. END OF REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE PERFORMING A TONSILLECTOMY, A TONSIL TIP IGNITED. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLASMABLADE TISSUE DISSECTION DEVICE GEI MEDTRONIC ADVANCED ENERGY LLC PLASMABLADE TNA 52727

Patients

Seq Age Sex Outcome Treatment
1 UNK NOT AVAILABLE