FDA Adverse Event Injury Summary report: N

LIFELINE PERSONAL RESPONSE SYSTEM

MDR report key: 2593054 · Received May 24, 2012

Report

Report Number
1220762-2012-00002
Event Type
Injury
Date Received
May 24, 2012
Date of Event
January 9, 2012
Report Date
May 15, 2012
Product Code
ILQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT DOES NOT APPEAR FROM AVAILABLE INFO, THERE WAS ANY MALFUNCTION OF THE DEVICE. OUR RECORDS DID NOT INDICATE A HELP CALL WAS INITIATED DURING THE TIME OF THE EVENT. PER PHILIPS LIFETIME AUTOALERT HELP BUTTON. INSTRUCTIONS FOR USE, (P/N 0940718 REV. 1) P. 4 STATES: "CAUTION: IN CERTAIN SITUATIONS, THE AUTOHELP BUTTON MAY NOT DETECT A FALL. A GRADUAL SLIDE FROM A SEATED POSITION SUCH AS FROM A WHEELCHAIR-MAY NOT REGISTER AS A FALL AND WOULD NOT BE DETECTED. IF YOU FALL AND NEED HELP, ALWAYS PUSH THE HELP BUTTON IF YOU ARE ABLE TO." THE PERSONAL HELP BUTTON HAD DETECTED OTHER FALLS FROM THIS SUBSCRIBER. IN ADDITION, THE BUTTON WAS DEPRESSED AFTER THE INCIDENT AND A HELP CALL WAS SUCCESSFULLY INITIATED. IN THE EVENT, THAT ADD'L INFO IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SENT.

Description of Event or Problem · 1

USER ALLEGED THAT DEVICE USER WAS FOUND ON HER SIDE ABOUT 5 FEET FROM HER RECLINER. SHE LANDED ON A HEATING GRATE WHICH HAD BENEATH IT A HOT WATER HEAT EXCHANGER THAT GENERATED 200 DEGREE TEMPERATURE. SHE LAID ON THE HEATING GATE UNCONSCIOUS FOR SEVERAL HOURS. SHE SUSTAINED BURNS ON HER ARMS AND SHOULDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFELINE PERSONAL RESPONSE SYSTEM SYSTEM, COMMUNICATION, POWERED ILQ AAHB

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization