FDA Adverse Event Malfunction Summary report: N

GCX

MDR report key: 258862 · Received January 14, 2000

Report

Report Number
258862
Event Type
Malfunction
Date Received
January 14, 2000
Date of Event
January 6, 2000
Report Date
January 7, 2000
Manufacturer
GCX CORPORATION
Product Code
IOY
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

GCX POLYMOUNT BRACKET SEPARATED WHEN SCREW FELL OUT AND A MONITOR WEIGHING APPROXIMATELY 55 POUNDS FELL. THE MONITOR WAS CAUGHT AND HELD BY A NURSE AND WAS ALSO HELD BY ONE ARM OF THE METAL BRACKET THAT WAS BENT BY THE WEIGHT OF THE MONITOR. THE SCREWS USED TO MOUNT THIS MONITOR WERE INSTALLED FLAT HEAD TOWARD THE BED. WHEN THE CAP STYLE BOLT ON TOP LOOSENS, THE SCREWS FALL OUT. THERE IS THE POTENTIAL TO CAUSE MAJOR PT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCX POLYMOUNT BRACKET IOY GCX CORPORATION * *

Patients

Seq Age Sex Outcome Treatment
1 * Other