FDA Adverse Event
Malfunction
Summary report: N
GCX
MDR report key: 258862
·
Received January 14, 2000
Report
- Report Number
- 258862
- Event Type
- Malfunction
- Date Received
- January 14, 2000
- Date of Event
- January 6, 2000
- Report Date
- January 7, 2000
- Manufacturer
- GCX CORPORATION
- Product Code
- IOY
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
GCX POLYMOUNT BRACKET SEPARATED WHEN SCREW FELL OUT AND A MONITOR WEIGHING APPROXIMATELY 55 POUNDS FELL. THE MONITOR WAS CAUGHT AND HELD BY A NURSE AND WAS ALSO HELD BY ONE ARM OF THE METAL BRACKET THAT WAS BENT BY THE WEIGHT OF THE MONITOR. THE SCREWS USED TO MOUNT THIS MONITOR WERE INSTALLED FLAT HEAD TOWARD THE BED. WHEN THE CAP STYLE BOLT ON TOP LOOSENS, THE SCREWS FALL OUT. THERE IS THE POTENTIAL TO CAUSE MAJOR PT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GCX | POLYMOUNT BRACKET | IOY | GCX CORPORATION | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |