FDA Adverse Event
Other
Summary report: N
KARL STORZ
MDR report key: 2588426
·
Received May 17, 2012
Report
- Report Number
- 9610617-2012-00016
- Event Type
- Other
- Date Received
- May 17, 2012
- Date of Event
- April 17, 2012
- Report Date
- May 15, 2012
- Manufacturer
- KARL STORZ GMBH & CO. KG
- Product Code
- KTG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE TIP OF THE KNIFE IS BROKEN OFF; IT IS APPROXIMATELY 7MM IN LENGTH. CONTACT AT HOSPITAL STATED THEY BELIEVED IT BROKE DUE TO THE DENSITY AND HARDNESS OF THE SEPTUM BONE. CONDITION OF INSTRUMENT IS CONSISTENT WITH DAMAGE CAUSED BY MECHANICAL OVERLOAD. THIS IS THE ONLY REPORT OF BREAKAGE OF THIS INSTRUMENT THAT WE HAVE RECEIVED.
Description of Event or Problem · 1
ALLEGEDLY, THE TIP OF THE SICKLE KNIFE BROKE OFF IN THE PT'S NASAL PASSAGE DURING AN ENT PROCEDURE; THE TIP FELL DOWN THE PT'S THROAT. THE DOCTOR PERFORMED A LARYNGOSCOPY AND RETRIEVED THE PIECE WITH A GRASPER. HE WENT ON TO COMPLETE ORIGINAL PROCEDURE; ONLY A MINOR DELAY CAUSED AND THERE WAS NO IMPACT ON PT. PT CONDITION POST OP WAS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KARL STORZ | SICKLE KNIFE | KTG | KARL STORZ GMBH & CO. KG | 628002 | BI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |