FDA Adverse Event Other Summary report: N

KARL STORZ

MDR report key: 2588426 · Received May 17, 2012

Report

Report Number
9610617-2012-00016
Event Type
Other
Date Received
May 17, 2012
Date of Event
April 17, 2012
Report Date
May 15, 2012
Manufacturer
KARL STORZ GMBH & CO. KG
Product Code
KTG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE TIP OF THE KNIFE IS BROKEN OFF; IT IS APPROXIMATELY 7MM IN LENGTH. CONTACT AT HOSPITAL STATED THEY BELIEVED IT BROKE DUE TO THE DENSITY AND HARDNESS OF THE SEPTUM BONE. CONDITION OF INSTRUMENT IS CONSISTENT WITH DAMAGE CAUSED BY MECHANICAL OVERLOAD. THIS IS THE ONLY REPORT OF BREAKAGE OF THIS INSTRUMENT THAT WE HAVE RECEIVED.

Description of Event or Problem · 1

ALLEGEDLY, THE TIP OF THE SICKLE KNIFE BROKE OFF IN THE PT'S NASAL PASSAGE DURING AN ENT PROCEDURE; THE TIP FELL DOWN THE PT'S THROAT. THE DOCTOR PERFORMED A LARYNGOSCOPY AND RETRIEVED THE PIECE WITH A GRASPER. HE WENT ON TO COMPLETE ORIGINAL PROCEDURE; ONLY A MINOR DELAY CAUSED AND THERE WAS NO IMPACT ON PT. PT CONDITION POST OP WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KARL STORZ SICKLE KNIFE KTG KARL STORZ GMBH & CO. KG 628002 BI

Patients

Seq Age Sex Outcome Treatment
1 Other