DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2012-00023
- Event Type
- Malfunction
- Date Received
- May 25, 2012
- Date of Event
- January 14, 2010
- Report Date
- February 3, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS FILED THE ORIGINAL MDR 2517506-2010-00010 ON 2010-02-01 AS CAUSE UNKNOWN. NEW INFORMATION WAS PROVIDED BY THE SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) ON (B)(6) 2010: ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HM CHROME CONTAMINATION. THE SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE ACCOUNT AND PERFORMED REPAIRS AND DE-CONTAMINATION PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT BASED ON A PRIOR SEQUENTIAL SAMPLE. THE SAMPLE WAS REPEATED AND A HIGHER RESULT WAS OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PERSCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | RXL MAX HM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |