FDA Adverse Event Malfunction Summary report: N

MACH1 GUIDE CATHETER

MDR report key: 2587379 · Received May 24, 2012

Report

Report Number
2134265-2012-03122
Event Type
Malfunction
Date Received
May 24, 2012
Date of Event
May 2, 2012
Report Date
May 2, 2012
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K020028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE BATCH NUMBER IS UNKNOWN; THEREFORE THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE CANNOT BE REVIEWED. IF ANY FURTHER RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS 2134265-2012-03121, 2134265-2012-03120, AND 2134265-2012-03114. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, WITHDRAWAL DIFFICULTIES OCCURRED. ACCESS WAS OBTAINED THROUGH RADIAL ARTERY. THE AORTIC ROOT WAS LARGE SO THE PHYSICIAN STARTED THE CASE WITH A MACH 1 VL 3, IT WAS TOO SMALL SO IT WAS EXCHANGED FOR A MACH 1 VL 3.5 AND THEN TO A MACH 1 VL 4.0 AND THEN A MACH 1 VL 4.5. DURING REMOVAL OF THE 3.5, 4.0 AND 4.5 DEVICES THE PHYSICIAN EXPERIENCED DIFFICULTY REMOVING THEM FROM THE SHEATH AT THE DISTAL 3MM PORTION (TIP). DURING REMOVAL OF THE 4.5 DEVICE, THE DEVICE GOT STUCK AND REQUIRED FORCE TO REMOVE IT. UPON REMOVAL A SMALL SHARD OF PLASTIC, APPROXIMATELY 3MM IN LENGTH, WAS OBSERVED ON THE WIRE AND HUB; THE PHYSICIAN DOESN'T KNOW IF IT CAME FROM THE SHEATH OR GUIDE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MACH1 GUIDE CATHETER CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE UNK497

Patients

Seq Age Sex Outcome Treatment
1 ARROW MEDICAL INTRODUCER SHEATH| MACH 1 VL3 GUIDE CATHETER| MACH 1 VL3.5 GUIDE CATHETER