FDA Adverse Event Malfunction Summary report: N

CORE UNIVERSAL DRIVER

MDR report key: 2584422 · Received May 18, 2012

Report

Report Number
1811755-2012-01868
Event Type
Malfunction
Date Received
May 18, 2012
Date of Event
April 27, 2012
Report Date
April 27, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CONTACT PINS IN THE POTTED TRIGGER ASSEMBLY WERE BURNT, AND THE GEARS ON THE FIRS STAGE DIDN'T ROTATE SMOOTHLY.

Description of Event or Problem · 1

THE CORE UNIVERSAL DRIVER WAS SENT FOR EVALUATION DUE TO OVERHEATING WHILE BEING TESTED. THERE WAS NO PT INVOLVEMENT, AND NO ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE UNIVERSAL DRIVER HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK