INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00281
- Event Type
- Malfunction
- Date Received
- May 21, 2012
- Date of Event
- April 11, 2012
- Report Date
- April 11, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- REFER TO H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT RT265 INFANT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR MISSING PORT PLUGS. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED BREATHING CIRCUIT. FURTHER INSPECTION REVEALED THAT THE PORT PLUG OF THE EXPIRATORY TUBE WAS NOT MISSING BUT RATHER LOOSE INSIDE THE PACKAGING. THE PORT PLUG OF THE DRY LINE HAD COME LOOSE AS WELL. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120125. CONCLUSION: WE WERE UNABLE TO DETERMINE THE ROOT CAUSE OF THE LOOSE PORT PLUGS. ALL PORT PLUGS ARE FULLY AND FIRMLY PRESSED INTO THE BREATHING CIRCUIT CONNECTORS BEFORE PACKAGING. MOREOVER, ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED PRIOR TO LEAVING THE PRODUCTION LINE, AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT IT IS LIKELY THAT THE PORT PLUGS BECAME LOOSE POST PRODUCTION, POSSIBLY DURING TRANSPORTATION OR STORAGE. WE HAVE NOT RECEIVED ANY OTHER COMPLAINTS OF THIS NATURE FOR THIS PRODUCT.
A HOSPITAL IN THE (B)(6) REPORTED THAT THE "WHITE CAP OF THE TUBING" OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT WAS MISSING. THIS WAS NOTICED PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT265 | 120125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |