FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2582335 · Received May 21, 2012

Report

Report Number
9611451-2012-00321
Event Type
Malfunction
Date Received
May 21, 2012
Report Date
April 27, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ONE OF THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: ONE OF THE COMPLAINT RT340 ADULT EVAQUA BREATHING CIRCUITS WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. THE EVAQUA EXPIRATORY TUBE WAS PRESSURE TESTED AND VISUALLY INSPECTED. RESULTS:THE PRESSURE TEST RESULT WAS OUTSIDE OF THE REQUIRED SPECIFICATION. FURTHER INSPECTION REVEALED THAT THE EXPIRATORY TUBE WAS PUNCTURED ABOUT 4CM FROM THE PATIENT END CONNECTOR. A LOT CHECK REVEALED THREE OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110615. CONCLUSION:BASED ON INSPECTION OF THE DAMAGE, IT IS LIKELY THAT THE EVAQUA EXPIRATORY TUBE WAS PUNCTURED OR SCRATCHED WITH A BLUNT OBJECT. ALL RT340 ADULT BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS DAMAGED POST-PRODUCTION. (B)(4). THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." "FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT FOUR RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS EACH HAD A HOLE IN THE EXPIRATORY LIMB. IT WAS FURTHER REPORTED THAN AN ERROR MESSAGE "LEAK" APPEARED ON THE EVITA 4 VENTILATOR DURING TESTING OF THE BREATHING CIRCUITS. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT FOUR RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS EACH HAD A HOLE IN THE EXPIRATORY LIMB. IT WAS FURTHER REPORTED THAT AN ERROR MESSAGE "LEAK" APPEARED ON THE EVITA 4 VENTILATOR DURING TESTING OF THE BREATHING CIRCUITS. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT340 110615

Patients

Seq Age Sex Outcome Treatment
1