V60 VENTILATOR

MDR report key: 2582086 · Received May 21, 2012

Report

Report Number
2031642-2012-00218
Event Type
Injury
Date Received
May 21, 2012
Date of Event
April 22, 2012
Report Date
April 23, 2012
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
MNT
PMA / PMN Number
K082660
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE DIRECTOR OF RESPIRATORY THERAPY (RT) REPORTED THE VENTILATOR ALARMED DUE TO OXYGEN NOT AVAILABLE WHILE TRANSPORTING THE PATIENT TO ICU VIA OXYGEN CYLINDER WITH REGULATOR. THE CUSTOMER REPORTED THE O2 TANK REGULATOR WAS OBSERVED TO BE FROZEN SOLID WHITE. LOSS OF THE OXYGEN SOURCE CAN BE DETRIMENTAL TO A PATIENT. THE PATIENT WAS BAGGED THE REMAINDER OF THE TRANSPORT. THE RT DIRECTOR REPORTED THE PATIENT WAS IN CRITICAL CONDITION AND DETERIORATING RAPIDLY AND WAS LATER INTUBATED AND DIED A FEW HOURS LATER BUT STATED THE PATIENT WAS TERMINAL AND THE PATIENT'S DEATH WAS NOT A RESULT OF THE ISSUE DURING TRANSPORT BUT A RESULT OF THEIR CONDITION. THE MANUFACTURER'S SERVICE TECHNICIAN TESTED THE DEVICE USING THE CUSTOMER CIRCUIT AND PARAMETER SETTINGS. THE SERVICE TECHNICIAN REPORTED THE REGULATOR THAT WAS IN USE ON THE O2 CYLINDER FELT COOL TO THE TOUCH. THE SERVICE TECHNICIAN REPORTED THE DEVICE STARTING ALARMING O2 NOT AVAILABLE WHEN THE REGULATOR DROPPED BELOW 100 PSI. THE SERVICE TECHNICIAN REPORTED HE SWITCHED OUT THE O2 TANK FOR A FULL ONE AND THE DEVICE STOPPED ALARMING. THE DEVICE PASSED FULL PERFORMANCE VERIFICATION TESTING. REVIEW OF THE DEVICE DIAGNOSTIC LOG NOTED O2 NOT AVAILABLE ALARM OCCURRENCES. THE MANUFACTURER'S CLINICAL SPECIALIST DISCUSSED THE REPORTED EVENT WITH THE RT DIRECTOR AND EXPLAINED THE POSSIBILITY OF A RAPID PRESSURE DROP CAUSING A DROP IN TEMPERATURE IN HIGH HUMIDITY ENVIRONMENTS (EITHER HUMIDITY PRESENT IN THE REGULATOR OR HUMIDITY IN THE ROOM). SINCE THE CUSTOMER HASN'T OBSERVED THIS DURING OTHER TRANSPORTS, THE SPOTLIGHT APPEARED TO BE ON THE O2 REGULATOR, O2 TANK FILLING-PROCEDURE AND/OR ENVIRONMENT. ALTHOUGH THERE IS A TEMPERATURE CHANGE AT THE REGULATOR DURING USE THE MOISTURE BUILDUP AND REPORTED FREEZING OF THE REGULATOR WAS NOT ASSOCIATED WITH THE DEVICE DIRECTLY. THE DEVICE HAS BEEN PLACED BACK INTO USE WITH NO FURTHER ISSUES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 V60 VENTILATOR VENTILATOR, CONTINUOUS MNT RESPIRONICS CALIFORNIA, INC. V60 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention